7.3 Cleaning the device
Personnel:
n
Operating personnel
WARNING!
Risk of cleaning agent entering the device
Electric shock
Only use a slightly damp cloth for cleaning.
Please also note the following:
n
Only use water and detergent to clean the control panel. Do not use
acetone or solvent. as these substances will permanently damage the
plastic surfaces.
7.4 Cleaning the air-cooled condenser
Personnel:
n
Operating personnel
1.
Switch off the device.
2.
Remove the front panel by grasping the bottom with both hands and
pulling the grille forward. Remove the front panel slowly and carefully
to avoid damage.
3.
Brush off or vacuum the condenser.
4.
Install the front panel again with care.
7.5 Check the heat transfer liquid
Contaminated heat transfer liquid must be replaced. Continued use of the
heat transfer liquid is only permitted following successful testing.
The heat transfer liquid must be tested as outlined in DIN 51529.
V07
Microcool