Table 20: Control parameters
ID Function
Unit
Command
38 Control parameter Xp
[–]
OUT_PAR_00_XX.X
40 Control parameter Tn (5 – 180 s; 181 = Off)
[s]
OUT_PAR_01_XXX
42 Control parameter Tv
[s]
OUT_PAR_02_XXX
44 Control parameter Td
[s]
OUT_PAR_03_XX.X
46 Control parameter KpE
[–]
OUT_PAR_04_XX.XX
48 Control parameter TnE (0 – 9000 s; 9001 = Off)
[s]
OUT_PAR_05_XXXX
50 Control parameter TvE (5 = Off)
[s]
OUT_PAR_06_XXXX
52 Control parameter TdE
[s]
OUT_PAR_07_XXXX.X
54 Correction limitation
[K]
OUT_PAR_09_XXX.X
56 Control parameter XpF
[–]
OUT_PAR_10_XX.X
60 Control parameter Prop_E
[K]
OUT_PAR_15_XXX
Table 21: Control
ID Function
Unit
Command
58 Set point offset
[K]
OUT_PAR_14_XXX.X
66 Control in control variable X: 0 = internal / 1 = external Pt /
2 = external analog / 3 = external serial / 5 = external Ethernet /
6 = external EtherCAT / 7 = external Pt 2 (only for Integral)
[–]
OUT_MODE_01_X
68 Offset source X for set point: 0 = normal / 1 = external Pt /
2 = external analog / 3 = external serial / 5 = external Ethernet /
6 = external EtherCAT / 7 = external Pt 2
[–]
OUT_MODE_04_X
Note (ID 66 and 68): If X = 3, the commands ID 66 and ID 68 cannot be executed in some constant temperature control
devices until an external temperature specification has been received (via the command ID 15).
Table 22: Rights
ID Function
Unit
Command
62 Master keyboard (equivalent to "KEY"): 0 = unlock / 1 = lock
[–]
OUT_MODE_00_X
64 Keyboard remote control unit (command): 0 = unlock / 1 = lock
[–]
OUT_MODE_03_X
Table 23: Status
ID Function
Unit
Command
74 Switch the device on / off (standby)
[–]
START / STOP
V1
Interface module LRZ 930