98 / 124
ECO Silver
E
Control parameters
The control parameters have been optimized at the factory for operation as a bath thermostat (with water as the heat
transfer liquid) with internal control. The standard parameters are already set as default also for the thermostatic
control of external applications with external control.
Depending on the application, the configuration can be adapted from case to case as required. Also the thermal
capacity and the viscosity of the heat transfer liquid affect the control behavior.
Note:
Only change the control parameters if you have adequate knowledge of control techniques.
If you have not connected any temperature sensor, read further here. For activated external control read (
The control corresponds to the set-point temperature with the current bath temperature and calculates the set value for
heating or cooling.
These control parameters can be set:
Description
Short form
Unit
Proportional range
Xp
K
Reset time
Tn
s
Derivative time
Tv
s
Damping
Td
s
If "
Tv manual/auto
" is set to "a" (automatic),
Tv
and
Td
cannot be changed. They are in this case derived from
Tn
with
fixed factors.
Consider the effect of the temperature limits Tih and Til (
7.4.5) on the control.
–
You activate the soft-key bar by pressing any key.
–
Access to the main menu level is obtained by pressing the enter
key .
–
Selection and confirmation of
Setup
Control
Contr.parameter
intern Pt100
.
The adjacent menu window appears. Apart from the control parameters
the currently set values are displayed.
–
Under the menu point "Tv man/auto" you can select between
manual and automatic entry using .
The selection is indicated in the menu line by a (automatic) or m (manual).
If "automatic" is selected, the entry is blocked for the parameters Tv and
Td.
–
Select and confirm parameters with or and .
The appropriate edit window appears with Min and Max figures for the
parameter values Xp, Tn, Tv and Td.
–
Change the value with or .
–
Single figures can be selected by pressing or .
–
Confirm your choice with the enter key .
–
By pressing (ESC) you are returned to the menu level without any change.
Xp
Min: 0.1
Max: 50.0
ESC
OK
- - -
12.0
ESC
OK
STOP
Xp
10.0
Tn
30
Tv man/auto
a
Tv
24 (a)
Td
4.0 (a)
E
.1
Internal control variable (internal temperature sensor)
Summary of Contents for ECO Silver E 4 S
Page 1: ...ECO Silver Operation manual...
Page 2: ......
Page 4: ......
Page 78: ...78 124 ECO Silver 10 2019 i...
Page 79: ...10 2019 i ECO Silver 79 124...
Page 80: ...80 124 ECO Silver 10 2019 i Appendix with settings...
Page 111: ...10 2019 i ECO Silver 111 124 5 Click on Continue installation 6 Click on the key Finish...
Page 113: ...10 2019 i ECO Silver 113 124...
Page 120: ...120 124 ECO Silver 10 2019 i Blank page...
Page 121: ...10 2019 i ECO Silver 121 124 Blank page...
Page 122: ...122 124 ECO Silver 10 2019 i...
Page 125: ......
Page 126: ......
Page 127: ......