30 / 124
ECO Silver
Connection of the cooling water
Note that the following conditions apply for the connection of the cooling water supply:
Cooling water pressure (feed - outlet)
max. 10 bar overpressure
Differential pressure (feed - outlet)
min. 3.0 bar
Cooling water temperature
10 to 15 °C recommended,
10 to 30 °C admissible with power restrictions)
Cooling water quantity
see Technical Data (
Cooling water hose for connection to the device
min. 13 mm
Ways of adjusting the pump flow
The circulation of the heat transfer liquid by the pump can be divided between internal (INT) and external (EXT) with the aid of
the selector switch at the front on the control head (flow distribution). The adjustment is continuously variable and is also possi-
ble at any time during operation. .
The adjustment between internal and external circulation is only practicable with a connected external consumer. A pump
connection set is needed to do this. This set is included as standard with cooling devices and with the heating devices E 4 S and
ET 15 S. With immersion thermostats and the other heating thermostats the pump connection set is available as an accessory
(
With a pure bath application the selector switch has to be set to INT.
Summary of Contents for ECO Silver E 4 S
Page 1: ...ECO Silver Operation manual...
Page 2: ......
Page 4: ......
Page 78: ...78 124 ECO Silver 10 2019 i...
Page 79: ...10 2019 i ECO Silver 79 124...
Page 80: ...80 124 ECO Silver 10 2019 i Appendix with settings...
Page 111: ...10 2019 i ECO Silver 111 124 5 Click on Continue installation 6 Click on the key Finish...
Page 113: ...10 2019 i ECO Silver 113 124...
Page 120: ...120 124 ECO Silver 10 2019 i Blank page...
Page 121: ...10 2019 i ECO Silver 121 124 Blank page...
Page 122: ...122 124 ECO Silver 10 2019 i...
Page 125: ......
Page 126: ......
Page 127: ......