![Lauda ECO Silver E 4 S Operation Manual Download Page 35](http://html1.mh-extra.com/html/lauda/eco-silver-e-4-s/eco-silver-e-4-s_operation-manual_4087659035.webp)
ECO Silver
35 / 124
Draining and changing the heat transfer liquid
–
Switch off the thermostat and withdraw the mains plug.
–
Allow the device and heat transfer liquid to cool down to or warm
up to room temperature.
–
Push a hose onto the bath drain point.
–
Drain the heat transfer liquid by opening the drain tap at the back
of the device.
1 Drain tap, cooling thermostats
2 Drain tap, heating thermostats
Completely drain the bath, external consumers, accessories and hose connections and flush or clean them (e.g. with new
heat transfer liquid).
Caution!
Contact with hot / cold heat transfer liquid
Scalds, frostbite
Bring heat transfer liquids to room temperature before
draining.
Make sure that the drain tap is closed after draining.
Caution!
Delay in boiling and thermal decomposition due to liquid residues
Burns, scalds, development of harmful vapors
Remove all old heat transfer liquid completely from the bath,
external consumers, accessories and hoses. Flush and clean
them with new heat transfer liquid.
1
2
Summary of Contents for ECO Silver E 4 S
Page 1: ...ECO Silver Operation manual...
Page 2: ......
Page 4: ......
Page 78: ...78 124 ECO Silver 10 2019 i...
Page 79: ...10 2019 i ECO Silver 79 124...
Page 80: ...80 124 ECO Silver 10 2019 i Appendix with settings...
Page 111: ...10 2019 i ECO Silver 111 124 5 Click on Continue installation 6 Click on the key Finish...
Page 113: ...10 2019 i ECO Silver 113 124...
Page 120: ...120 124 ECO Silver 10 2019 i Blank page...
Page 121: ...10 2019 i ECO Silver 121 124 Blank page...
Page 122: ...122 124 ECO Silver 10 2019 i...
Page 125: ......
Page 126: ......
Page 127: ......