ECO Silver
It is possible to apply an offset value to the temperature, which is provided by an external temperature sensor and to
process it as the set value.
The bath temperature can therefore be operated, for example, -15 °C below the temperature of a reactor measured by
the external temperature sensor.
–
You activate the soft-key bar by pressing any key.
–
Access to the main menu level is obtained by pressing the enter
key .
–
Selection and confirmation of
Setup
Control
Setpoint offset
.
Offset source
.
The adjacent menu window appears.
–
Select and confirm offset source with or and .
–
The setpoint offset is deactivated with "off".
–
The entry window appears on selecting the menu point
Diff.set/actual.
The minimum and maximum possible offset values and the current offset
value are displayed.
–
Change the value with or .
–
Single figures can be selected by pressing or .
–
By pressing (+/-) the arithmetic sign can be changed.
–
Confirm your choice with the enter key .
–
By pressing (ESC) you are returned to the menu level without any change.
Offset value
Min: -500.0
Max: 500.0
ESC
OK
+ / -
-15.0
ESC
OK
STOP
off
extern Pt100
extern USB
C
.
3
Setpoint offset operating mode (Diff.set/actual)
Summary of Contents for ECO Silver E 4 S
Page 1: ...ECO Silver Operation manual...
Page 2: ......
Page 4: ......
Page 78: ...78 124 ECO Silver 10 2019 i...
Page 79: ...10 2019 i ECO Silver 79 124...
Page 80: ...80 124 ECO Silver 10 2019 i Appendix with settings...
Page 111: ...10 2019 i ECO Silver 111 124 5 Click on Continue installation 6 Click on the key Finish...
Page 113: ...10 2019 i ECO Silver 113 124...
Page 120: ...120 124 ECO Silver 10 2019 i Blank page...
Page 121: ...10 2019 i ECO Silver 121 124 Blank page...
Page 122: ...122 124 ECO Silver 10 2019 i...
Page 125: ......
Page 126: ......
Page 127: ......