3-10
(E)
BL55-RU
8
Move the scale, and make sure that the amplitudes A and B (see Fig. 3-14) of the Lissajous’ figure are
within the range of the specified values over the entire length.
<Specified values>
A/B signal output type ..... Output format F : Amplitude 0.7 Vp-p to 1.3 Vp-p
Output format G : Amplitude 0.8 Vp-p to 1.2 Vp-p
Analog output type ........... Output format H : Amplitude 0.3 Vp-p to 0.6 Vp-p
The center of the Lissajous’ figure is 2.5 V.
If the reference value outputs are not obtained or the center of the Lissajous’ figure is not at 2.5 V, refer
to the attached “Attaching the Slider” (see page 3-6).
Fig. 3-14
Lissajous
B
A