6-2
Performing Measurements
LxT Manual
Use of a Windscreen
Wind blowing across the microphone generates pressure
fluctuations on the microphone diaphragm that can produce
errors in the measurement. As a result, when performing
measurements in the presence of low level airflows, it is
recommended that a windscreen be placed over the
microphone. Larson Davis provides the WS001 windscreen,
a 3 1/2” diameter ball made of open cell foam which can be
placed over the microphone and preamplifier as shown in
FIGURE 6-1.
FIGURE 6-1 Position of Windscreen
Performing Measurements
This section describes the steps for performing basic sound
level measurements.
Starting the Measurement
The LxT uses a single range for
sound level measurements, so there
is no need to select a range. as part
of making a measurement.
The
Live
tab displays current acoustic data that is not being
recorded or stored. Pressing the
9
(
RUN/PAUSE) key
causes the LxT to begin storing data, which is displayed on
the
Overall
tab.
Summary of Contents for SoundExpert LxT
Page 1: ...LxT Manual for SoundTrack LxT SoundExpert LxT TM...
Page 55: ...4 18 Measurement Settings Tabs LxT Manual...
Page 81: ...5 26 View Spectrum Normalized Optional LxT Manual...
Page 93: ...6 12 Data Storage After Improper Shutdown LxT Manual...
Page 110: ...LxT Manual Certification 7 17...
Page 143: ...11 6 Control Panel Data Explorer LxT Manual...
Page 167: ...12 24 Options LxT Manual...
Page 179: ...14 4 User LxT Manual FIGURE 14 5 shows the User tab FIGURE 14 5 User Tab...
Page 185: ...15 6 File System LxT Manual...
Page 191: ...16 6 Miscellaneous Parameters LxT Manual...
Page 206: ...LxT Manual Technical Specifications A 9 Typical Z Weight Frequency Response...