LxT Manual
Preferences
12-9
FIGURE 12-11 shows the
Preferences
tab.
FIGURE 12-11 Preferences Tab
To change the fields on the
Preferences
tab, navigate and
select values as described in previous sections.
Microphone Correction
When using a free-field microphone, a correction can be
applied to provide a random incidence response or, when
using a random incidence microphone, a correction can be
applied to provide a free-field response. Highlight the
Mic
Corr.
field and press
5
to open the Microphone Correction
menu, as shown in FIGURE 12-12.
FIGURE 12-12 Microphone Correction Menu
To correct a random incidence microphone to obtain a free-
field response, select
RI -> FF
and press
5
.
To correct a free-field microphone to obtain a random
incidence response, select
FF -> RI
and press
5
.
The default mode is
Off
.
Auto-Store
The LxT provides three Auto-Store options to enhance your
data gathering activities:
Summary of Contents for SoundExpert LxT
Page 1: ...LxT Manual for SoundTrack LxT SoundExpert LxT TM...
Page 55: ...4 18 Measurement Settings Tabs LxT Manual...
Page 81: ...5 26 View Spectrum Normalized Optional LxT Manual...
Page 93: ...6 12 Data Storage After Improper Shutdown LxT Manual...
Page 110: ...LxT Manual Certification 7 17...
Page 143: ...11 6 Control Panel Data Explorer LxT Manual...
Page 167: ...12 24 Options LxT Manual...
Page 179: ...14 4 User LxT Manual FIGURE 14 5 shows the User tab FIGURE 14 5 User Tab...
Page 185: ...15 6 File System LxT Manual...
Page 191: ...16 6 Miscellaneous Parameters LxT Manual...
Page 206: ...LxT Manual Technical Specifications A 9 Typical Z Weight Frequency Response...