C
-
30
Glossary and Formulas
LxT Manual
16.61 for an exchange rate of 5dB (L
OSHA
)
20.00 for an exchange rate of 6dB (L
Avg
)
n is the total number of samples taken in the measurement.
The sample rate is 32 samples per second.
Time Weighting
The response speed of the detector in a sound level meter.
There are several speeds used.
Slow
: The time constant is 1 second (1000 ms). This is the
slowest and is commonly used in environmental noise
measurements.
Fast
: The time constant is 1/8 second (125 ms). This is a less
commonly used weighting but will detect changes in sound
level more rapidly.
Impulse
: The time constant is 35ms for the rise and 1.5
seconds (1500 ms) for the decay. The reason for the double
constant is to allow the very short signal to be captured and
displayed.
Wavelength (l)
The distance between peaks of a propagating wave with a
well defined frequency. It is related to the frequency through
the following equation
where
c
is the sound speed and
f
is the frequency in Hz. It
has the dimensions of length.
c
f
--
=
Summary of Contents for SoundExpert LxT
Page 1: ...LxT Manual for SoundTrack LxT SoundExpert LxT TM...
Page 55: ...4 18 Measurement Settings Tabs LxT Manual...
Page 81: ...5 26 View Spectrum Normalized Optional LxT Manual...
Page 93: ...6 12 Data Storage After Improper Shutdown LxT Manual...
Page 110: ...LxT Manual Certification 7 17...
Page 143: ...11 6 Control Panel Data Explorer LxT Manual...
Page 167: ...12 24 Options LxT Manual...
Page 179: ...14 4 User LxT Manual FIGURE 14 5 shows the User tab FIGURE 14 5 User Tab...
Page 185: ...15 6 File System LxT Manual...
Page 191: ...16 6 Miscellaneous Parameters LxT Manual...
Page 206: ...LxT Manual Technical Specifications A 9 Typical Z Weight Frequency Response...