P1202 / P1301 L-ESD set
Pulse level
(HV)
Pulse current
(
i
P
)
Input voltage oscilloscope (50
Ω)
using different attenuators
[kV]
[A]
-
[V]
10 dB
[V]
20 dB
[V]
30 dB
[V]
0.5
12
0.6
0.19
0.06
0.02
1
24
1.2
0.38
0.12
0.04
1.5
36
1.8
0.56
0.18
0.06
2
48
2.4
0.75
0.24
0.08
2.5
60
3
0.94
0.3
0.09
3
72
3.6
1.13
0.36
0.11
3.5
84
4.2
1.31
0.42
0.13
4
96
4.8
1.5
0.48
0.15
4.5
108
5.4
1.69
0.54
0.17
5
120
6
1.88
0.6
0.19
5.5
132
6.6
2.06
0.66
0.21
6
144
7.2
2.25
0.72
0.23
Table 12: Input voltage present at oscilloscope using different attenuators
The P1202 field source's measurement output can be used to calculate the magnetic flux density
B
generated by the field source at the position of the test IC.
The voltage equivalent of the pulse current i
p
has to be measured at the field
source’s
measurement output for this purpose. The respective field quantities can then be calculated using
the conversion factor defined in Table 13.
B = K1 ∙ i
P
i
P
= K4 ∙ u
out
Table 13: Conversion factor of the P1202 field source to calculate the field quantities