8 - 35
LANGER
EMV-Technik
DE-01728 Bannewitz
[email protected]
www.langer-emv.com
P1202 / P1301 L-ESD set
Technical Parameters
4
P1202 L-ESD
– General Parameters
4.1
Dimensions (h/w/d)
(180 x 96 x 96) mm
Weight
700 g
Pulse parameter
Shape
0.2 / 2.5 ns
Frequency
0.1 Hz
– 10 Hz
Max. current
± 144 A
Voltage
± (0.1
– 6) kV
Ammeter
Measurement output
50
Ω, SMB
Shunt
0.1
Ω
Current correction factor R
-26 dB
Ω
Connector
– input
50
Ω Fischer (D103A023)
Table 1: P1202 general parameters
P1202 L-ESD
– Magnetic Flux Density B
4.2
𝐻𝑉
in kV
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
5.5
6
V
iout
in V
0.6
1.2
1.8
2.4
3
3.6
4.2
4.8
5.4
6
6.6
7.2
𝑖
P
in A
12
24
36
48
60
72
84
96
108
120
132
144
D70 h03 spacer, height h = 3 mm
B in µT
0.293 0.586 0.878 1.17
1.46
1.76
2.05 2.34 2.64
2.93
3.22
3.51
D70 h10 spacer, height h = 10 mm
B in µT
0.149 0.298 0.446 0.595 0.744 0.893 1.04 1.19 1.34
1.49
1.63
1.79
Table 2: Magnetic flux density generated depending on the high voltage
𝐻𝑉
and probe current
𝑖
P
P1202 L-ESD - Shunt
4.3
Bandwidth
3 GHz
Measurement error
approx. 10 %
Attenuator settings
x 20 A / V
→ 26 dB
Deskew B to
𝑣
out
at measurement output
1 ns
Table 3: P1202 shunt parameters