24 - 35
LANGER
EMV-Technik
DE-01728 Bannewitz
[email protected]
www.langer-emv.com
P1202 / P1301 L-ESD set
Set-up of the P1202 / P1301 L-ESD Set
7
shows the basic set-up for immunity measurements with the P1202 field source as an
example. The P1202 / P1301 L-ESD set, the ICE1 set IC test environment
4
as well as different
measuring and test devices
5
are required for the measurement set-up:
Figure 14: Set-up for immunity measurements using the ICE1 set and P1202 ESD magnetic field source.
All devices which are not included in the scope of delivery are marked with an asterisk (*).
The following components, not included in the scope of delivery, are also needed:
-
ICE1 set,
IC test environment (Langer EMV-Technik GmbH)
-
Measuring and test devices
(e. g. Oscilloscope)
-
Test board
for the test IC
The test board together with the test IC is inserted into the GND 25 ground plane of the
measurement set-up. The field source (here P1202) is located above the test IC by using a spacer
ring. The BPS 203 is used to supply the high voltage and control the field source.
An oscilloscope and PC are needed to monitor the device under test and/or control the
measurement set-up.
Only the P1202 features a measurement output, which allows controlling generated pulse current.
4
not included in the scope of delivery
5
not included in the scope of delivery