8
Options
8.2.2
Factory Setting of the Digital Outputs
The LSB module for digital outputs 1
...
4 is activated at the factory. In case of a later installa-
tion or exchange, the
LSB address 03
must be set on the module using only 2 rotary switches.
The LSB module for digital outputs 5
...
8 is activated at the factory. In case of a later installa-
tion or exchange, the
LSB address 51
must be set on the module using only 2 rotary switches.
Digital output 1
Term. 13/14
Locked faults
Digital output 2
Term. 23/24
Warning
Digital output 3
Term. 33/34
Limit value 1 (exceeding of 400
ppm
CO
e
, 60
s)
Digital output 4
Term. 43/44
Limit value 2 (undershooting of 0.5 vol.
%
O
2
, 30
s)
Digital output 5
Term. 13/14
Calibration
Digital output 6
Term. 23/24
Cold start
Digital output 7
Term. 33/34
Measure
Digital output 8
Term. 43/44
Warning