Maintenance
Table 5:Test gas composition
5.1.11
Checking/Calibrating with Reference Measurement
•
The system must be in operation (ideally at the desired operating point).
•
Perform the calibration as described in Chapters
5.1.2 Checking/Calibrating the O
5.1.9.3 Perform a Calibration of the CO/H
Test gas
Composition
O
2
[Vol.%]
CO
e
[ppm]
N
2
[Vol.%]
A
21
0
Rest
B
3
0
Rest
C
3
300
Rest
*
CO Equivalent CO
e
is the sum of all components in the exhaust gas. In test gases, it is represented by CO and H
2
in pro-
portion of 2:1, e.g. 300
ppm
CO
e
= 200
ppm CO +100
ppm
H
2
.
**
In the test gas device, the probe can be calibrated/tested with oil-free pressure air or synthetic air from test gas bottle. In
a CO
e
-free and defined surrounding, the probe can also be tested/calibrated outside of the test gas device.