12
Other
Settings
About… lists the Serial Number, Firmware Version, Bluetooth MAC address,
Bluetooth Version and Test Count. The Test Count shows the number of
complete tests that have been performed over the lifetime of the meter.
Tap
to return to the Settings menu.
Ranges Enabled allows the option of having test results that are out of the
range of the reagent system displayed in red. The default setting is on.
Select Disk Detection to display an error message and abort the test when
improper testing conditions, such as No Disk, No Disk Cover, Used Disk or
Meter Check Disk, are present that would result in inaccurate or no test
results. The default setting is OFF. Tap
to return to the Settings menu.
RANGES
Freshwater FF104 (Code 4351)
Test Factor
Range
Display Abbreviation
Alkalinity
0-250 ppm
ALK
Ammonia
0.0-4.0 ppm
AMMO
Hardness
0-500 ppm
G HARD
Nitrate
0-250 ppm
NITRATE
Nitrite
0.0-2.0 ppm
NITRITE
pH
4.5-10.0
pH
Phosphate
0.0 - 2.0 ppm
PHOS
Saltwater FF203 (Code 4352)
Test Factor
Range
Display Abbreviation
Alkalinity
0-350 ppm
ALK
Ammonia
0.0-4.0 ppm
AMMO
Calcium
200-800 ppm
Ca
Magnesium
500-2200 ppm
Mg
Nitrate
0-60 ppm
NITRATE
Nitrite
0.0-2.0 ppm
NITRITE
pH
6.5-10.0
pH
Phosphate
0.0 - 2.0 ppm
PHOS
Test results that are out of range of the reagent system wil be
RED
.
RED
test results may
not be accurate. Go SETTINGS>Other Settings to turn the Ranges Enabled feature OFF.
Testing samples for Nitrate immediately after treatment with a chlorine neutralizer
containing sodium thiosulfate will give low results. Retest in 2-3 days.
If the concentration for one test factor is signifi cantly out of the range for the reagent
system, the accuracy of the results for the other test factors may be aff ected.