Ozone
, Indigo
COD3 Plus Colorimeter 10.16
PROCEDURE–LOW RANGE
A. PREPARATION OF HR REAGENT
NOTE: The quantity of Indigo Blue Stock Solution (3989) supplied will prepare one batch
of HR Reagent for the High Range Ozone procedure or fi ve batches of HR Reagent for
the Low Range Ozone procedure.
1. Use the 50 mL graduated cylinder to carefully add 45 mL of *Ozone Buff er (3991)
to amber glass bottle marked HR Reagent (3988-MT-G).
2. Use the 5 mL transfer pipet (0329) and pump (30527) to add 5 mL of Indigo Blue
Stock Solution (3989) to the amber glass bottle. Cap and mix.
B. DETERMINATION OF OZONE
3. Use the 1.0 mL transfer pipet (2-2170) and pump (30527) to add 1.0 mL of HR
Reagent (3988) to each of 2 clean tubes (0290).
4. If chlorine is present add 3 drops Chlorine Inhibitor (3990) to each tube. Cap
tubes.
5. Take one of the prepared tubes (0290) and sampling apparatus (0681) to sampling
site.
6. Lower end of tubing of sampling apparatus to desired depth. Slowly withdraw and
depress plunger several times to purge syringe and tubing. Slowly withdraw plunger
to fi ll purged syringe.
7. Remove plastic tubing from syringe. Remove cap from the prepared tube. Place tip
of syringe against inside of the prepared tube. Slowly depress plunger and fi ll to the
10 mL line and cap. This is the Sample Tube.
NOTE: DO NOT SHAKE OR INVERT THE SAMPLE.
8. Fill the second prepared tube (0290) to the 10 mL line with ozone free water. This is
the Reagent Blank.
9. Press and hold
until colorimeter turns on.
10. Press
ENTER
to select
Testing Menu
.
11. Select
All Tests
(or another sequence containing
071 Ozone IB LR
) from
Testing
Menu
.
12. Scroll to and select
071 Ozone IB LR
from menu.
13. Insert the Reagent Blank tube into chamber, close lid and select
Scan Blank
.
14. Insert reacted Sample Tube into chamber, close lid and select
Scan Sample
. Record
result.
15. Press
to turn colorimeter off or press
EXIT
to exit to a previous menu or
make another menu selection.
Testing
Summary of Contents for COD3 Plus
Page 2: ......
Page 29: ...COD3 Plus Colorimeter Test Procedures Firmware Version 2 10 and greater ...
Page 30: ......
Page 38: ...Ammonia Nitrogen Low Range COD3 Plus Colorimeter 10 16 Testing ...
Page 42: ...Ammonia Nitrogen High Range COD3 Plus Colorimeter 10 16 Testing ...
Page 46: ...Boron COD3 Plus Colorimeter 10 16 Testing ...
Page 72: ...Dissolved Oxygen COD3 Plus Colorimeter 10 16 Testing ...
Page 76: ...Fluoride COD3 Plus Colorimeter 10 16 Testing ...
Page 90: ...Phosphate Low Range COD3 Plus Colorimeter 10 16 Testing ...
Page 98: ...Silica Low Range COD3 Plus Colorimeter 10 16 Testing ...
Page 102: ...Silica High Range COD3 Plus Colorimeter 10 16 Testing ...
Page 108: ...Testing Sulfide Low Range COD3 Plus Colorimeter 10 16 ...
Page 114: ...Testing Turbidity COD3 Plus Colorimeter 10 16 ...