Cobalt
COD3 Plus Colorimeter 10.16
PROCEDURE
1. Press and hold
until colorimeter turns on.
2. Press
ENTER
to select
Testing Menu
.
3. Select
All Tests
(or another sequence containing
029 Cobalt
) from
Testing
Menu
.
4. Scroll to and select
029 Cobalt
from menu.
5. Rinse a tube (0290) with sample water. Fill to 10 mL with sample.
6. Insert the tube into chamber, close lid and select
Scan Blank
.
7. Remove the tube from colorimeter.
8. Use the 1.0 mL pipet (0354) to add 1 mL of *Cobalt Buff er (4852). Cap and
mix.
9. Use the other 1.0 mL pipet (0354) to add 1 mL of *Cobalt Indicator Reagent
(4853). Cap and mix.
10. Wait 3 minutes.
11. Use the 0.5 mL pipet (0353) to add 0.5 mL Stabilizer Solution (4854). Cap and
invert 15 times to thoroughly mix.
12. Wait 5 minutes. DO NOT MIX.
13. Insert tube into chamber, close lid and select
Scan Sample
. Record result in ppm
cobalt.
14. Press
to turn the colorimeter off or press
EXIT
to exit to a previous menu
or make another menu selection.
NOTE: For best possible results, a reagent blank should be determined to account for
any contribution to the test result by the reagent system. To determine the reagent
blank, follow the above test procedure to scan a distilled or deionized water blank.
Then follow the above procedure to perform the test on a distilled or deionized water
sample. This test result is the reagent blank. Subtract the reagent blank from all
subsequent test results of unknown samples. It is necessary to determine the reagent
blank only when a new lot number of reagents is obtained.
Testing
Summary of Contents for COD3 Plus
Page 2: ......
Page 29: ...COD3 Plus Colorimeter Test Procedures Firmware Version 2 10 and greater ...
Page 30: ......
Page 38: ...Ammonia Nitrogen Low Range COD3 Plus Colorimeter 10 16 Testing ...
Page 42: ...Ammonia Nitrogen High Range COD3 Plus Colorimeter 10 16 Testing ...
Page 46: ...Boron COD3 Plus Colorimeter 10 16 Testing ...
Page 72: ...Dissolved Oxygen COD3 Plus Colorimeter 10 16 Testing ...
Page 76: ...Fluoride COD3 Plus Colorimeter 10 16 Testing ...
Page 90: ...Phosphate Low Range COD3 Plus Colorimeter 10 16 Testing ...
Page 98: ...Silica Low Range COD3 Plus Colorimeter 10 16 Testing ...
Page 102: ...Silica High Range COD3 Plus Colorimeter 10 16 Testing ...
Page 108: ...Testing Sulfide Low Range COD3 Plus Colorimeter 10 16 ...
Page 114: ...Testing Turbidity COD3 Plus Colorimeter 10 16 ...