COD3 Plus Colorimeter 10.18
17
COMPUTER CONNECTION
OUTPUT
USB
COMPUTER CONNECTION
USB Type A, USB mini B, Order Cable Code 1720.
WATERLINK CONNECT 2
The meter may be interfaced with any Windows-based 64 bit computer by using
the LaMotte WaterLink Connect 2 program and a USB cable. The program will store
test information and results in a local database, and allow for exporting this data to
a comma separated value (CSV) fi le. The meter will send the following data: time/
date stamp, name of test, sample value, sample units, meter name, and location.
To download WaterLink 2 go to http://www.waterlinkconnect.com/ . Then select
“Products” > “WaterLink Connect 2 Application” > “Download”.
BATTERY
BATTERY/AC OPERATION
The colorimeter may be operated on battery power, using a USB wall adapter or USB
computer connection. If using the meter as a bench top unit, use the wall adapter if
possible to extend the battery life. The meter will remain on when the USB adapter is
used.
To charge the battery with the wall adapter, plug the smaller end of the USB cable
(USB mini B connector) into the meter and the larger end of the USB cable (USB Type
A connector) into the wall adapter. Plug the wall adapter into an AC outlet. Reinsert the
USB port plug after charging.
To charge the battery from a computer, plug the smaller end of the USB cable (USB
mini B connector) into the meter and the larger end of the USB cable (USB Type A
connector) into a USB port on a computer.
The battery icon will show no bars and fl ash when the unit fi rst turns on. Then the
indicator will indicate the battery status by showing 0, 1, 2, 3 or 4 bars.
It will take 5 hours to fully charge a low battery. The battery icon will fl ash when the
battery is charging. The battery icon will show four bars and stop fl ashing when it is
fully charged. The charging circuit will automatically switch to a fl oat charge when the
battery is fully charged. The charger may remain connected. Some computers will NOT
supply power to their USB ports during standby operation. The wall adapter will charge
the unit continuously.
The battery icon will show no bars and continuously fl ash if the battery is getting low
but the unit will still operate normally. A “Low Battery” message on the status bar of
Summary of Contents for COD3 Plus
Page 2: ......
Page 29: ...COD3 Plus Colorimeter Test Procedures Firmware Version 2 10 and greater ...
Page 30: ......
Page 38: ...Ammonia Nitrogen Low Range COD3 Plus Colorimeter 10 16 Testing ...
Page 42: ...Ammonia Nitrogen High Range COD3 Plus Colorimeter 10 16 Testing ...
Page 46: ...Boron COD3 Plus Colorimeter 10 16 Testing ...
Page 72: ...Dissolved Oxygen COD3 Plus Colorimeter 10 16 Testing ...
Page 76: ...Fluoride COD3 Plus Colorimeter 10 16 Testing ...
Page 90: ...Phosphate Low Range COD3 Plus Colorimeter 10 16 Testing ...
Page 98: ...Silica Low Range COD3 Plus Colorimeter 10 16 Testing ...
Page 102: ...Silica High Range COD3 Plus Colorimeter 10 16 Testing ...
Page 108: ...Testing Sulfide Low Range COD3 Plus Colorimeter 10 16 ...
Page 114: ...Testing Turbidity COD3 Plus Colorimeter 10 16 ...