Hydrazine
COD3 Plus Colorimeter 10.16
PROCEDURE
1. Press and hold
until colorimeter turns on.
2. Press
ENTER
to select
Testing Menu
.
3. Select
All Tests
(or another sequence containing
049 Hydrazine
) from
Testing
Menu
.
4. Scroll to and select
049 Hydrazine
from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with sample.
6. Insert tube into chamber, close lid and select
Scan Blank
.
7. Remove tube from colorimeter. Use the 1 mL pipet (0354) to add 4 mL of
*Hydrazine Reagent A (4841). Cap and mix.
8. Use the 0.15 g spoon (0727) to add one measure of Hydrazine Reagent B Powder
(4842). Cap and shake vigorously for 10 seconds. Wait 2 minutes for maximum
color development. An undissolved portion of Hydrazine Reagent B may remain in
bottom of tube without adversely aff ecting results.
9. At the end of the 2 minute waiting period, mix, insert tube into chamber, close lid
and select
Scan Sample
. Record result.
10. Press
to turn colorimeter off or press
EXIT
to exit to a previous menu or
make another menu selection.
NOTE: For best possible results, a reagent blank should be determined to account for
any contribution to the test result by the reagent system. To determine the reagent
blank, follow the above test procedure to scan a distilled or deionized water blank. Then
follow the above procedure to perform the test on a distilled or deionized water sample.
This test result is the reagent blank. Subtract the reagent blank from all subsequent
test results of unknown samples. It is necessary to determine the reagent blank only
when a new lot number of reagents is obtained.
Testing
Summary of Contents for COD3 Plus
Page 2: ......
Page 29: ...COD3 Plus Colorimeter Test Procedures Firmware Version 2 10 and greater ...
Page 30: ......
Page 38: ...Ammonia Nitrogen Low Range COD3 Plus Colorimeter 10 16 Testing ...
Page 42: ...Ammonia Nitrogen High Range COD3 Plus Colorimeter 10 16 Testing ...
Page 46: ...Boron COD3 Plus Colorimeter 10 16 Testing ...
Page 72: ...Dissolved Oxygen COD3 Plus Colorimeter 10 16 Testing ...
Page 76: ...Fluoride COD3 Plus Colorimeter 10 16 Testing ...
Page 90: ...Phosphate Low Range COD3 Plus Colorimeter 10 16 Testing ...
Page 98: ...Silica Low Range COD3 Plus Colorimeter 10 16 Testing ...
Page 102: ...Silica High Range COD3 Plus Colorimeter 10 16 Testing ...
Page 108: ...Testing Sulfide Low Range COD3 Plus Colorimeter 10 16 ...
Page 114: ...Testing Turbidity COD3 Plus Colorimeter 10 16 ...