APPLICATION:
Drinking and surface waters, domestic and industrial waste
water.
RANGE:
0.00–3.00 ppm Zinc
MDL:
0.05 ppm
METHOD:
Zinc forms a blue colored complex with Zincon in a solution
buff ered at pH 9.0. Other heavy metals are complexed by
cyanide and the zinc cyanide complex is released by the
addition of formaldehyde before the other metal cyanide
complexes are destroyed. Sodium ascorbate is added to
reduce the interference of manganese.
SAMPLE HANDLING &
PRESERVATION:
Sample should be analyzed within 6 hours after collection.
The addition of hydrochloric acid will help preserve the
metal ion content, however the acid should be neutralized
before analysis.
INTERFERENCES:
The following ions interfere in concentrations greater than
those listed.
Ion
mg/L
Ion
mg/L
Cd(II)
1
Cr(III)
10
Al (III)
5
Ni(II)
20
Mn (II)
5
Co (II)
30
Fe (III)
7
CrO4(II)
50
Fe (II)
9
Testing
Zinc
COD3 Plus Colorimeter 10.16
Summary of Contents for COD3 Plus
Page 2: ......
Page 29: ...COD3 Plus Colorimeter Test Procedures Firmware Version 2 10 and greater ...
Page 30: ......
Page 38: ...Ammonia Nitrogen Low Range COD3 Plus Colorimeter 10 16 Testing ...
Page 42: ...Ammonia Nitrogen High Range COD3 Plus Colorimeter 10 16 Testing ...
Page 46: ...Boron COD3 Plus Colorimeter 10 16 Testing ...
Page 72: ...Dissolved Oxygen COD3 Plus Colorimeter 10 16 Testing ...
Page 76: ...Fluoride COD3 Plus Colorimeter 10 16 Testing ...
Page 90: ...Phosphate Low Range COD3 Plus Colorimeter 10 16 Testing ...
Page 98: ...Silica Low Range COD3 Plus Colorimeter 10 16 Testing ...
Page 102: ...Silica High Range COD3 Plus Colorimeter 10 16 Testing ...
Page 108: ...Testing Sulfide Low Range COD3 Plus Colorimeter 10 16 ...
Page 114: ...Testing Turbidity COD3 Plus Colorimeter 10 16 ...