Manuel Technique OsmsoeurRO400C
Version 3
Page 18 / 26
Control functions
The water purifier is equipped with some control functions to facilitate the diagnostic in case of trouble.
Displaying of the functioning times
Close the tap for purified water – Press for 3 sec. “Calibration” – Release for 1 sec. – Press again “Calibration” for 1 sec.
The water purifier cannot show a
number of functioning hours lower than
10 hours
Displaying of shutdowns for too low pressure
When the 3 LEDs have switched off at the end of displaying of the
functioning times, press for 1 sec. “Calibration” button.
1 flash = 1 shutdown
Count how many times the LED flashes (max. 255)
Reset the meters to zero
Reset also to zero the meter for the
replacing f filter 2.
•
Press and keep pressed "Calibration" button,
•
The diode for filter 2 lights up, goes out and
then lights up again,
•
Release the button.