![Lakeshore 336 User Manual Download Page 20](http://html1.mh-extra.com/html/lakeshore/336/336_user-manual_3369259020.webp)
8
c
HAPTER
1:
Introduction
Model 336 Temperature Controller
1.3.2 Sensor Input
Configuration
1.3.3 Thermometry
Thermocouple
option
Model 3060
Sensor
Tempera-
ture Coeffi-
cient
Input Range
Excitation
Current
Display
Resolution
Measurement
Resolution
Electronic
Accuracy
(at 25 °C)
Measurement Temperature
Coefficient
Electronic
Control
Stability
13
Thermocouple
3060
Positive
±50 mV
NA
0.1 µV
0.4µV
±1 µV ±0.05%
of rdg
12
(0.1 µV + 0.001% of rdg)/°C
±0.8µV
13
Control stability of the electronics only, in ideal thermal system
TABLE 1-4
Thermocouple option input specifications
Capacitance
option
Model 3061
Sensor
Tempera-
ture Coeffi-
cient
Input Range
Excitation
Current
Display
Resolution
Measurement
Resolution
Electronic
Accuracy
(at 25 °C)
Measurement Temperature
Coefficient
Electronic
Control
Stability
14
Capacitance
3061
Positive or
negative
0.1 nF to 15 nF
3.496 kHz 1 mA
square wave
0.1 pF
0.05 pF
±50 pF ±0.1%
of rdg
2.5 pF/°C
0.1 pF
1 nF to 150 nF
3.496 kHz 10 mA
square wave
1 pF
0.5 pF
±50 pF ±0.1%
of rdg
5 pF/°C
1 pF
14
Control stability of the electronics only, in ideal thermal system
TABLE 1-5
Capacitance option input specifications
Diode/RTD
Thermocouple
Measurement
type
4-lead differential
2-lead differential, room
temperature compensated
Excitation
Constant current with
current reversal for RTDs
NA
Supported sensors
Diodes
: Silicon, GaAlAs
RTDs
: 100
)
Platinum (option), 1000 ) Plat-
inum, Germanium,
Carbon-Glass, Cernox™, and Rox™
Most thermocouple types
Standard curves
DT-470, DT-670, DT-500-D, DT-500-E1,
PT-100, PT-1000, RX-102A, RX-202A
Type E, Type K, Type T, AuFe
0.07% vs. Cr, AuFe 0.03% vs. CR
Input connector
6-pin DIN
Screw terminals in a ceramic
isothermal block
TABLE 1-6
Sensor input configuration
Number of inputs
4 (8 with Model 3062)
Input configuration
Inputs can be configured from the front panel to accept any of the
supported input types. Thermocouple and capacitance inputs require
an optional input card that can be installed in the field.
Supported option cards
Thermocouple (3060), capacitance (3061), or scanner (3062)
Option slots
1
Isolation
Sensor inputs optically isolated from other circuits but not each other
A/D resolution
24-bit
Input accuracy
Sensor dependent, refer to Input Specifications table
Measurement resolution
Sensor dependent, refer to Input Specifications table
Maximum update rate
10 rdg/s on each input, 5 rdg/s when configured as 100 k
)
NTC RTD
with reversal on
Maximum update rate (scanner)
The maximum update rate for a scanned input is 10 rdg/s distributed among
the enabled channels. Any channel configured as 100 k
)
RTD with reversal on
changes the update rate for the channel to 5 rdg/s
Autorange
Automatically selects appropriate NTC RTD or PTC RTD range
User curves
Room for 39 200-point CalCurves™ or user curves
SoftCal™
Improves accuracy of DT-470 diode to ±0.25 K from 30 K to 375 K; improves
accuracy of platinum RTDs to ±0.25 K from 70 K to 325 K; stored as user curves
Math
Maximum and minimum
Filter
Averages 2 to 64 input readings
Summary of Contents for 336
Page 4: ...Model 336 Temperature Controller...
Page 6: ...Model 336 Temperature Controller...
Page 26: ...14 cHAPTER 1 Introduction Model 336 Temperature Controller...
Page 54: ...42 cHAPTER 3 Installation Model 336 Temperature Controller...
Page 84: ...72 cHAPTER 4 Operation Model 336 Temperature Controller...
Page 104: ...92 cHAPTER 5 Advanced Operation Model 336 Temperature Controller...
Page 164: ...152 cHAPTER 7 Options and Accessories Model 336 Temperature Controller...
Page 178: ...166 cHAPTER 8 Service Model 336 Temperature Controller...