62
c
HAPTER
3:
Installation
Model 372 AC Resistance Bridge ands Temperature Controller
3.5.7 Temperature
Sensors
Resistive temperature sensors should be treated as any other resistor being mea-
sured. One difference is that four lead devices are normally calibrated with a specific
arrangement of current and voltage leads. Follow all manufacturer recommenda-
tions on installation, including lead arrangement, or errors in temperature measure-
ment will result.
FIGURE 3-7
Example of instrument grounding
Model
372
Model
372
case to earth
ground through
the power cord
Shield tied to
scanner connector
(no gap in shielding)
Short cables
(shielded only)
Long cables
(shielded and guarded)
Dewar is connected
to earth ground
Scanner case is electrically
connected to the Dewar
case (connects isolated
circuits to earth ground)