174
c
HAPTER
6:
Computer Interface Operation
Model 372 AC Resistance Bridge and Temperature Controller
RDGK?
Kelvin Reading Query
Input
RDGK? <input/channel>[term]
Format
nn
<input/channel>
Specifies which input or channel to query: A (control input),
1 to 16 (measurement input).
Returned
<kelvin value>[term]
Format
±nnn.nnnE±nn
Remarks
Also see RDGST? If no curve is present for the channel, the returned reading will be
zero.
RDGPWR?
Excitation Power Reading Query
Input
RDGPWR? <input/channel>[term]
Format
nn
<input/channel>
Specifies which input or channel to query: A control input),
1 to 16 (measurement input).
Returned
<excitation power>[term]
Format
±nnn.nnnE±nn
Remarks
Also see RDGST? Returns the most recent power calculation for the specified channel.
RDGR?
Resistance Reading Query
Input
RDGR? <input/channel>[term]
Format
nn
<input/channel>
Specifies which input or channel to query: A (control input),
1 to 16 (measurement input).
Returned
<ohm value>[term]
Format
±nnn.nnnE±nn
Remarks
Also see the RDGST? command.
RDGST?
Input Reading Status Query
Input
RDGST? <input/channel>[term]
Format
nn
<input/channel>
Specifies which input or channel to query: A (control input),
1 – 16 (measurement input).
Returned
<status bit weighting>[term]
Format
nnn
Remarks
The integer returned represents the sum of the bit weighting of the status flag bits. A
“000” response indicates a valid reading is present. If the channel being queried is not
the actively scanning channel, the status will represent the last state prior to the
channel change.
Bit
Bit Weighting
Status Indicator
0
1
CS OVL
1
2
VCM OVL
2
4
VMIX OVL
3
8
VDIF OVL
4
16
R. OVER
5
32
R. UNDER
6
64
T. OVER
7
128
T. UNDER