6.6.1 Interface Commands
175
|
www.lakeshore.com
RDGSTL?
Reading Settle Status Query
Input
FILTERST? [term]
Returned
<Control input status>,<Active Measurement input Channel Status>[term]
Format
n,n
<status>
0 = Valid, 1 = Firmware Filter Settling, 2 = Hardware Settling
RELAY
Relay Parameter Command
Input
RELAY <relay number>,<mode>,<channel alarm>,<alarm
type>[term]
Format
n,n,nn,n
<relay number>
Specifies which relay to configure: 1 or 2.
<mode>
Specifies relay mode. 0 = off, 1 = on, 2 = Alarms, 3 = Sample
heater zone, 4=Warmup heater zone.
<channel alarm>
Specifies which input or channel alarm activates the relay
when the relay is in alarm mode: A (control input), 1 to 16
(measurement input), 0 = scan channel.
<alarm type>
Specifies the channel alarm type that activates the relay
when the relay is in alarm mode. 0 = Low alarm, 1 = High
Alarm, 2 = Both Alarms.
Example
RELAY 1,2,2,0[term]
—relay 1 is setup in Alarms mode and activates when channel 2
low alarm activates.
RELAY?
Relay Parameter Query
Input
RELAY? <relay number>[term]
Format
n
<relay number>
Specifies which relay to query: 1 or 2.
Returned
<mode>,<channel alarm>,<alarm type>[term]
Format
n,nn,n (refer to command for description)
RELAYST?
Relay Status Query
Input
RELAYST? <relay number>[term]
Format
n
<relay number>
Specifies which relay to query: 1 or 2.
Returned
<status>[term]
Format
n
0 = off, 1 = on.
SCAN
Scanner Parameter Command
Input
SCAN<input/channel>,<autoscan>[term]
Format
nn,n
<input/channel>
Specifies which input or channel to switch the scanner to:
1 to 16 (measurement input) (or 1 to 8 if a Model 3708
scanner is installed).
<autoscan>
0 = autoscan feature off, 1 = autoscan feature on
Example
SCAN 5,1[term]
— change the scanner to channel 5 and begin autoscanning.
Remarks
This command is only active if an optional scanner is installed.
SCAN?
Scanner Parameter Query
Input
SCAN?[term]
Returned
<input/channel>,<autoscan>[term]
Format
nn,n (refer to command for description).