Setup for Simulation
Manikin Setup
Cleaning the Fluid and Blood
System
After using the bleeding and or fluid system, the system must be
cleaned to prevent residue from clogging the valves and tubes. Keep
any wounds used connected while flushing the system with a cleaning
solution.
1 Ensure the manikin power is on.
2 Ensure the manikin’s internal reservoirs are drained. See procedure
for emptying.
3 Fill a bottle with a 60% Isopropanol solution.
4 Connect the bottle containing the alcohol solution to either the
blood or fluid fill connector and air connector to fill the manikin
with cleaning solution. Perform this action for both reservoirs.
CLEANING THE FLUID SYSTEM
1 From the Instructors application, open <Circulation and fluids>
tab, check the Secretions box for urine, and then select polyuria.
Check the Secretions boxes for Sweat, Ears, Eyes, Nose and
Mouth.
2 Allow the Fluid system to flush with the cleaning solution until the
internal reservoir is empty.
3 When finished, uncheck all of the Secretions boxes.
CLEANING THE BLOOD SYSTEM
1 From the <Instructor Application>, open Circulation and fluids
tab, check the boxes for Upper port and Lower port, then select
Venous from the adjacent drop-down menus. Move the sliders
(to the right) to maximum bleeding rates.
2 Allow the Bleeding system to flush with the cleaning solution until
the internal reservoir is empty.
3 The blood system should be flushed repeatedly until the
discharged fluids are clear.
4 When finished, uncheck all of the boxes (in the Instructors
application).