LABORIE Urocap III Owner’s Manual
UC3-UM01
34
Control Panel Commands
Suggested Button
Label
Definition
Type
Comments
First Sensation
First Urge
Capacity
Valsalva
Cough
Area Start
Area End
First Sensation
First Urge
Capacity
Valsalva
Cough
AS:Pves:0
AE:
Event
Event -
Unique
Marks an event with the definition
text as Event Annotation.
An event that happens only once
and will replace any existing
recorded event.
Uroflow
Full path of CFG file
Example: c:\cfg\
Config
Opens a .cfg file.
Auto Start
N/A
Auto Start
Sets up Uroflow Auto Start
Set Zeroes
N/A
Set Zeroes
Sets all channels to zero.
Pves = Pura or
Pura = Pves
Channel X = Channel Y
Example:
Pves = Pura or Pura =
Pves
Equalize
Sets the left channel value to the
right.
Patient Info
N/A
Patient Info
Opens Patient Info window.
File Saving Directory
The file saving directory option allows you to select the storage directory/folder for saving the UDS data files.
To change the Storage Directory:
1.
Click
Options
on the menu bar and select
File Saving Directory
.
Result
: The
Change Data Directory
window appears
2.
Click
Browse
and select a new directory.
3.
Click
OK
to close the Browse for Folder window.
4.
Click the
Create Dir.
button to set the new directory.
5.
Click
OK
to close the Change Data Directory window.
Auto Mark Peaks
The Mark Peaks feature automatically marks peak events for all channels when the test is stopped.
1.
Click Options > Auto Mark Peaks
to enable Mark Peaks.
Result
: After running and stopping a test, a peak event for each channel is added.
2.
Click
Options
> Auto
Mark Peaks
to disable Mark Peaks.