background image

This Way Up – Indicates correct upright position of the
transport package.

has met European Union health, safety, and environmental requirements,
which ensure consumer safety

Fragile – Contents of the transport package are fragile
and therefore shall be handled with care

(Conti...)

Keep Dry – Transport package shall be kept away from rain.

The CE mark (an acronym for the French "Conformite Europeenne")
certifies that a product

Year of Manufacture Used on PRODUCT DATA

Prima DNT Trainer

5

Issue 1.0

6167500-795

PRIMA DNT Trainer

Printed on May 2013

Summary of Contents for PRIMA DNT Trainer

Page 1: ...ent as well as to avoid injury while operating instrument understanding this manual completely before use is highly recommended User Manual PRIMA DNT Trainer Operating Surgical Microscopy R Part No 6167500 795 Issue1 0 Printed on May 2013 ...

Page 2: ...N AND TESTING MUST BE CONDUCTED TO ENSURE CONTINUED SAFE USE OF THIS INSTRUMENT WARNING TO AVOID RISK OF ELECTRIC SHOCK THIS EQUIPMENT MUST ONLY BE CONNECTED TO A SUPPLY MAIN WITH PROTECTIVE EARTH OR DAMAGE TO THIS INSTRUMENT AND OR INJURY TO THE OPERATOR OR PATIENT MAY OCCUR WARNING ENSURE THAT THE VOLTAGE APPLIED TO THE UNIT IS THE SAME AS THE VOLTAGE THAT IS INDICATED ON THE DATA PLATE OR DAMAG...

Page 3: ...F THIS INSTRUMENT AS DAMAGE TO THE UNIT MAY OCCUR SEE MAINTENANCE SECTION FOR DETAILED CLEANING INSTRUCTION CAUTION MEDICAL ELECTRONIC EQUIPMENT NEEDS SPECIAL PRECAUTIONS REGARDING EMC AND NEEDS TO BE INSTALLED AND PUT INTO SERVICE ACCORDING TO THE EMC INFORMATION PROVIDED IN THE ACCOMPANYING DOCUMENTS CAUTION PORTABLE AND MOBILE RF COMMUNICATIONS EQUIPMENT CAN AFFECT MEDICAL ELECTRICAL EQUIPMENT ...

Page 4: ...of Parts 7 System Diagram 8 9 Installation Instruction 10 11 12 12 13 14 15 15 16 Cleaning Servicing 17 Ambient Requirement 18 Troubleshooting Table 19 20 Disposal 21 Specification 22 Dimensions 23 Installation Electric Connection Controls Instruction for using the microscope Changing the Objectives Eyepieces Replacing the Illumination source Adjustment of tension Issue 1 0 ...

Page 5: ...ticular surgery from five different magnifications 3 Cold light illumination with a high intensity 50W LED lamp is provided using a fiber optic guide for proper illumination The illumination is further adjustable up to its most suitable brightness using intensity control knob suitably located at the suspension arm and is easily approachable to the surgeon 4 When the microscope is not in use the su...

Page 6: ... exclamation mark as shown here Please pay special attention to these safety notes Safety is only ensures when this instrument is operated properly Please read this manual carefully before turning the instrument on Also read through the user s manuals on the other equipment used with this instrument You may obtain further information from our service organization or authorization representative Th...

Page 7: ...medical imaging systems and equipment are unpredictable It is cautionary to mention that mobile phones or radios not be used in the proximity of the equipment Anything concerning repairs customization and upgradation this instruction must be carried out by LABOMED authorized representative failing which LABOMED product warranty will not be applicable The instrument must be operated for its intende...

Page 8: ...n site The power cord being used in the one designed for use with this instrument Make sure that all Requirement for operation are fulfilled Go through the checklist Re attach or close any covers panels or caps which have been removed or opened Pay special attention to warning symbols on the instrument triangular warning signs with exclamation marks labels and any parts such as screws or surfaces ...

Page 9: ...evice Directive 93 42 EEC REF Catalog Number Protective Earth 5 Issue 1 0 of 7 5 kg Please consult the user manual for the suspension system about the admissible maximum load safety compliance brand name Accompanying Document must be consulted ...

Page 10: ...which ensure consumer safety Fragile Contents of the transport package are fragile and therefore shall be handled with care Conti Keep Dry Transport package shall be kept away from rain The CE mark an acronym for the French Conformite Europeenne certifies that a product Year of Manufacture Used on PRODUCT DATA Prima DNT Trainer 5 Issue 1 0 ...

Page 11: ... 2 Column 3 Swivel arm and Suspension arm assembly with fibre optic cable 4 eVALUX Illumination Box 5 Inclined coupling with magni changer assembly and objective as ordered 6 Observation Head inclined or ergo as ordered 7 Paired Eye Pieces as ordered 8 Power Cord 9 Set of sterilizable caps 10 Installation Kit a Allen Wrench 5 00mm b Allen Wrench 8 00mm 11 Instruction cum Operating Manual Service M...

Page 12: ... 10 11 Magnichanger Locking Knob 12 Common Main Objective 13 Carrier Locking Knob 14 Swivel Arm 15 Handles 16 Filter Knob 17 Binocular head locking knob Suspension arm hydraulic movement lock Illumination Control Knob Binocular head with eyepieces Magnichanger Knob Fig 1 5 Prima DNT Trainer 7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 Issue 1 0 ...

Page 13: ...STEM DIAGRAM 6 Prima DNT Trainer 6133300 801 6133400 801 6132012 6132012 6135100 800 6121035 8 6122015 6133200 801 6133250 801 6167500 815 6167500 803 6167500 500 6137300 6137000 000 Issue 1 0 6165000 979 ...

Page 14: ...piece 6 6132010 10x Eyepiece 7 6121035 45 inclined Binocular Head 8 6122015 210 Straight binocular tube 9 6133250 801 250mm Objective 10 6133200 801 200 mm Objective 11 6133300 801 300mm Objective 12 6133400 801 400mm Objective 13 6137300 Table Mount 14 6135100 800 iVU S5 Digital Camera Module 15 6165000 979 Indian power cord 16 6165000 974 Shuko Power cord 17 6165000 973 US Power cord 18 6162086 ...

Page 15: ...oosening of swivel arm lock knob A Install the swivel arm assembly on the upright column 1 figure 3 Tighten the screw 2 as shown in fig 3 for proper securing the swivel arm with suspension arm 7 3 Retrieve the eVALUX illumination box with small column assembly from the packing box Fix it in between the swivel and suspension arm assembly as shown in Fig 4 and secure it by tighten the screws provide...

Page 16: ...the light guide has been routed in such a way that the carrier system and the surgical microscope are not obstructed and that they can be moved in their entire range of movement without stretching extreme kinking or twisting of the light guide 7 5 Install the binocular head and eyepieces on the magni changer Secure the binocular head with head locking screw shown as A in figure 6a The light guide ...

Page 17: ...level before switching on the system It is located in front of the suspension arm shown as A in Fig 8 Brightness of field of view can be adjusted as per user comfort using intensity control knob To save burning life of LED switch OFF the appliance if the microscope is not in use for longer time 9 2 Intensity control knob 9 4 Camera Potentiometer output in center shown as D in fig 8 9 5 Camera outp...

Page 18: ...er IPD scale according to your convenience Ref Fig 10 1 Adjust to highest magnification with one of the rotating knobs 10a provided at magnification changer 2 Fine focusing is done through knob 10b provided on objective as shown in fig 10 3 Absolute centering of observation area in field of view can be done by manual handle 10c 4 Make sure that the magnification changer is engaged in the index poi...

Page 19: ...e microscope image displays optimum contrast and excellent detail recognition along with a large depth of field The bright microscope image is a particular benefit in vitreoretinal surgery The objective lenses with fine focusing knob and focal lengths of 250 mm to 400 mm are available for different working distances Main microscope Prima DNT Trainer Fig 11 11 The standard equipment includes eyepie...

Page 20: ...mination box as shown in fig 13 Detach the fibre optic cable and replace the illumination assembly A with new assembly Secure back the illumination box with screws Fuse replacement 14 The fuse is located with the AC inlet provided on the back side of illumination box as shown in fig 14 Use a flat head screw driver to open the fuse compartment Two fuses are provided in this i e one is live fuse and...

Page 21: ... safe working condition thru sufficient cooling facility provided with proper free and forced air circulation by the fans provided in electrical box Further instrument is designed with a inbuilt safety mechanism with auto thermal cut off if the temperature of LED is above 70 C In case if thermal cut off fails no risk will happen to the instrument only LED may get fuse Here user needs to replace th...

Page 22: ... distilled water to which a dash of household dish washing liquid has been added Wipe the surface only with a damp under no circumstances with a wet clothe Any remaining marks can be easily removed using the following aides For thorough cleaning of optical surfaces use the optics cleaning set or damp optics cleaning wipes available from specialized dealers Remove minor contamination such as dust s...

Page 23: ...tation and storage Temperature Rel humidity without condensation Air pressure 40 C 70 C 10 100 500hPa 1 060hPa Ambient Requirement The unit meets the essential requirements stipulated in Annex I to the 93 42 EEC directive governing medical devices The unit is marked with and is compliance to IEC 60601 1 2007 3rd edition CE 18 Issue 1 0 ...

Page 24: ...medy TROUBLESHOOTING TABLE 18 Surgical field illumination inoperative Light guide not properly inserted in microscope arm Insert light guide as far as it will go Failure of electronics Illumination surgical field using an OR illuminator Contact service dept Switch off via microswitch on suspension system Move suspension arm of suspension system into working position 19 Issue 1 0 This instrument is...

Page 25: ... Illuminate surgical field using an Or illuminator Contact service dept Up Down motion Friction adjustment knob Loosen friction adjustment of knob on on suspension system knob on suspension system suspension system tightened too firmly as require too stiff No image visible in Magnichanger is not Index magnichanger properly field of view indexed properly Problem Possible Cause Remedy 20 Issue 1 0 ...

Page 26: ...DISPOSAL 19 Prima DNT Trainer 21 Issue 1 0 Disposal must comply with locally applicable laws and regulations ...

Page 27: ...tive f 250mm manual fine focus Light source 50W LED Built in filters Green Yellow Vertical movement of arm 500mm Microscope carriers Straight Carrier Accessories iVu S5 Beam splitter integrated digital camera Module or 45 Type Table Mount Pillar Height 400 mm Weight of complete microscope 22 Kg Approx Elevation Stroke 500 mm System Height in Horizontal Position 713 mm Prima DNT Trainer 22 Issue 1 ...

Page 28: ...Dimensions 21 DIMENSIONS Prima DNT Trainer 23 Table Mount 600 mm 265 mm 230mm 400 mm 71 3 mm 250m m 135 150 1 3 5 90 1 4 5 Issue 1 0 ...

Page 29: ...ce LABOMED and Prima DNT TRAINER are registered trademarks of Labo America Inc With a policy of continuous development Labo America Inc reserves the right to change design and specifications without prior notice Labo America Inc 920 Auburn Court Fremont CA 94538 U S A Telephone 510 445 1257 Fax 510 991 9862 sales laboamerica com A16372 LABOMED ISO 13485 ...

Reviews: