Prima DNT Trainer
7.4 Retrieve the carrier assembly from the packing
and follow as below refer Fig. 5.
Install the coupling to the suspension arm by
sliding the guiding shaft (1) to the suspension
arm.
Lock the inclined coupling with the threaded plug
(3).
Make sure that safety screw must tight in place
at position (2) to avoid sudden falling of the
coupling. Safely screw tagged within the arm.
7.5 Mounting the light guide
Switch off the illumination system
Make sure that the light guide has been routed in
such a way that the carrier system and the
surgical microscope are not obstructed, and that
they can be moved in their entire range of
movement without stretching, extreme, kinking or
twisting of the light guide.
7.5 Install the binocular head and eyepieces on the
magni-changer. Secure the binocular head with
head locking screw shown as (A) in figure 6a.
The light guide comes pre-routed through
eVALUX illumination box (that houses the LED
illumination system) and the suspension arm.
Insert the light guide into the receptacle in the
microscope till it clicks into position as shown as
A in Fig. 6
Fig. 6 a
Fig. 5
Fig. 6
11
2
1
3
A
A
Issue 1.0
6167500-795
PRIMA DNT Trainer
Printed on May 2013