Prima DNT Trainer
17.1 Cleaning of optical surfaces:
17.2
Caution:
17.3
Cleaning mechanical surfaces:
17.4 Sterilization
The MAXlite multi-layer coating of the optical components (e.g. eyepieces objective
lenses) ensured optimum image quality.
Image quality is impaired by even slight contamination. To protect the internal optics from
dust, the system should never be left without the objective lens, binocular tube and
eyepieces. After use, cover the system to protect it from dust, Always store optical
components and accessories which are not being used in dust-free cases.
Clean the external surfaces of optical components as required.
Never spray fumigates directly on the instruments or optical surfaces.
Do not use any chemical detergents or aggressive substances. These may damage the
optical surfaces.
Remove coarse dirt (splashes or blood etc.) using distilled water to which a dash of
household dish-washing liquid has been added. Wipe the surface only with a damp,
under no circumstances with a wet clothe. Any remaining marks can be easily removed
using the following aides.
For thorough cleaning of optical surfaces, use the optics cleaning set or damp optics
cleaning wipes(available from specialized dealers).
Remove minor contamination such as dust, streaks, etc. using a clean microfiber cleaning
cloth(available from specialized dealer).
All mechanical surface of the equipment can be cleaned by wiping them with a damp
cloth . Do not use any aggressive or abrasive detergents.
Wipe off any residue with a mixture of 50% ethyl alcohol and 50% distilled water plus a
dash of household dish-washing liquid.
The asepsis sets available from LABOMED contain rubber caps, sleeves and grips which
can be sterilized in autoclaves. We recommend the following program for sterilization:
Sterilization temperature:
134° C
Sterilizing time:
10 minute
Instrument :
Standard, Autoclove
CLEANING AND SERVICING
17
17
Issue 1.0
6167500-795
PRIMA DNT Trainer
Printed on May 2013