Magna
MAGNA
15. USING THE MICROSCOPE
WARNING: INSTRUMENT IS UNSAFE FOR MRI ENVIRONMENTS.
6129000-795
Issue 1.7
Printed on May,2020
A
A.1
B
C
D
E
42
F
G
Fig. 19
Fig. 20
CAUTION:
Upon Installation, ABA System may feel stuck due to longtime
Inoperative in transit or storage. This temporary mild locking may
happen due to material Property of ABA braking Material.
To overcome this situation,-Switch on the microscope, Press Push
Button with one hand and Push ABA System with other hand to
dislodge this temporary Air lock. Repeat it 7-8 times, ABA System
will start working normally.
SETTING UP THE MICROSCOPE:
a.
For stability of microscope, lock all four brakes on the
base caster wheels.
b.
Optimize the Auto Balancing Arm (ABA) position by using
clutch motions shown as (A) and (A.1) to get the proper
field of vision as shown in Fig. 19.
c.
Desired working distance can be obtained by rotating
the ring shown as (B) on the variable objective lens.
Fine focusing can be done by the same ring as
shown in Fig. 19.
d.
After focusing, lock swivel arm, suspension arm and ABA
arm movements using lock knobs shown as (C) as
shown in Fig. 19.
e.
Adjust the inter- pupillary distance (IPD) of the eyepieces
using the IPD adjustment knob shown as (D) in Fig.
19.
f.
The illumination is controlled through the intensity control
knob as shown as (E). Rotate it clockwise or anticlockwise
to achieve the desired illumination level as shown in
Fig. 19.
SETTING UP THE MAGNIFICATION:
a.
Use the knobs (F) provided on the Magnichanger to
select the desired magnification level.
b.
Rotate these clockwise to increase and Anticlockwise to
decrease the level of magnification as shown in Fig. 20.
K