Magna
MAGNA
14.2 ELECTRICAL CONNECTIONS
6129000-795
Issue 1.7
Printed on May,2020
1
2
3
A
B
C
37
Fig. 11
Fig. 12
Fig. 13
Connect the power cable to the AC inlet socket (1) located
at the back of the swivel arm as shown in Fig. 11.
Switch on the power using the ON/OFF switch (2).
Note: The power supply is designed with universal input
100V-240V AC, 50/60Hz. To plug in the power supply,
follow the instruction on the product label (3) located at
the bottom of the arm as shown in fig.- 11.
1. REPLACING THE ILLUMINATION SOURCE
a. Remove the swivel arm covers by unscrewing the screws
in top and bottom as shown in Fig. 12.
b. Remove the connections on the power supply and replace
the illumination assembly (A) in Fig. 12 with a new assembly
and refix the connections.
c. Assemble back the arm covers.
2. REPLACING THE FUSE
a. Locate the fuse at the AC inlet (B) as shown in Fig.13.
b. Use a flat head screw driver to open the fuse compar-
tment (C). Two fuses are provided: one is a live fuse
and the second is a spare fuse.
c. Replace the blown fuse with the spare fuse and close the
fuse compartment.
Note: For proper ratings of Fuse used in microscope, refer
to the label shown in fig.- 11.