Magna
Magna is intended for use in the electromagnetic environment specified below. The customer or
user of the Magna should ensure that it is used in such an environment.
Guidance and Manufacturer’s Declaration
Electromagnetic Emissions
All Equipment and Systems
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
Immunity Test
Level
Compliance Level
Electromagnetic Environment
Guidance
ESD IEC
61000-4-2
± 2,4,8kv contact
±2,4,8,15kv Air
± 2,4,8kv Contact
±2,4,8,15kv Air
EFT IEC 61000-4-4
AC Mains or DC Input ±2
Kv, 100 kHz PRF. I/O
Ports ±1 Kv, 100 kHz PRF
AC Mains or DC Input ±2 Kv,
100 kHz PRF. I/O Ports ±1 Kv,
100 kHz PRF
Surge IEC
61000-4-5
AC Mains, Line to
Ground ± 0.5,1,2Kv
Ac Mains, Line ±0.5,1Kv
DC Input (>3m), Line to
Ground ±0.5,1,2Kv
DC Input (>3m), Line to
Line ±0.5,1Kv
I/O, Line to Ground ±2
Kv (Outdoor lines only)
AC Mains, Line to Ground ±
0.5,1,2Kv
Ac Mains, Line ±0.5,1Kv
DC Input (>3m), Line to
Ground ±0.5,1,2Kv
DC Input (>3m), Line to Line
±0.5,1Kv
I/O, Line to Ground ±2 Kv
(Outdoor lines only)
VOLTAGE DIPS,
DROPOUTS, AND
INTERRUPTIONS, IEC
61000-4-11
Voltage Dips
(<16A)
Power Frequency
50/60Hz Magnetic
Field IEC 61000-4-8
30A/m
30A/m
6129000-795
Issue 1.7
Printed on May,2020
57
MAGNA
GUIDANCE TABLES
(continued)
Voltage
Interruptions (all
input current)
100% drop, 0.5 periods,
0°,45°,90°,135°,180°,225°,
270° ,315°
100% dip, 1 period
30% dip, 25/30 periods
100% drop, 5 seconds
100% drop, 0.5 periods,
0°,45°,90°,135°,180°,225°,
270° ,315°
100% dip, 1 period
30% dip, 25/30 periods
100% drop, 5 seconds
IEC 60601 Test
Power frequency magnetic field should be
that of a typical commercial or hospital
Environment.
Main power quality should be that of a
typical commercial or hospital environment.
Main power quality should be that of a
typical commercial or hospital environment.
Floors should be wood, concrete or ceramic
tile. If floors are synthetic, the R/H should be
at least 30%.
Main power quality should be that of a
typical commercial or hospital Environ-
ment. If the user of the Prima Magna
requires continued operation during
power mains interruption, it is recommen-
ded that the Prima Magna be powered
from an uninterruptable power supply or
battery.