BA_072_2020-02_10.00 Pressure transmitter PASCAL Ci4
Page 15/28
6.4 The Menue Tree
In the following, the display and adjustment options are described by their position in the
menue tree. An overview of the menue tree can be found on the last page of this docu-
ment.
6.4.1 Main menue
The main menue has the following entries:
Menue entry
Description
Quick setup
Selection of the most important settings
Adjustment
Adjustment functions for pressure measurement and current out-
put
Display
Functions for configuring the display
Measurement/output
Configuration of the pressure measurement and current output
Diagnosis
Diagnostic information such as min/max values
Simulation
Simulation of pressure and current for function check of the
measurement chain
Communication
Information and settings regarding HART communication
System
Device data and function such as factory reset
Tabelle 6: The main menue
6.4.2 "Quick Setup" menue
In Quick Setup, basic configuration options are combined to make it possible to quickly
configure key functions. All functions of the quick setup can also be found at another posi-
tion in the menue tree.
The following functions are available in the Quick Setup menue:
Menue entry
Description
Sprache/Language
Select menue language
Unit Pressure
Select unit for measured pressure (see 6.4.4.1)
Unit Temperature
Select unit for sensor temperature and for ATC-temperature (if
applicable) (see 6.4.4.1)
Lower Range Value
Setting of the pressure value that should correspond to 4 mA
(start of range) (see 6.4.5.4)
Upper Range Value
Setting of the pressure value that should correspond to 20 mA
(end of range) (see 6.4.5.4)
Damping
Setting the damping of the output signal (see 6.4.5.1)
Device ID
Setting the device ID (see 6.4.9.1)
Config.-memory
Configuration memory: reading, writing and status (see 6.4.9.3)
Table 7: "Quick Setup" menue