LaboGene ScanSpeed 1248 Instruction Manual Download Page 7

07052013RS01

Cat. No. 9.900.900.758

CopyRight © 2013

Page 

7

of 

20

1-2.1.1.  Safety Precautions

Before using the instrument, please read this operation manual to ensure correct usage through understanding. Incorrect 
handling  of  the  instrument  could  possibly  result  in  personal  injury  or  physical  damage  on  the  instrument  or  its 
accessories.   

1.

 

ALWAYS  locate  the  instrument  on  a  flat,  rigid  and  stable  table  capable  of  withstanding  the  weight  of  the 
instrument and its spinning operation.   

2.

ALWAYS  make  a  safety  zone  of  30  cm  around  the  centrifuge  to  indicate  that  neither  hazardous  materials  nor 
persons should be permitted within the area during operation. 

 

ALWAYS  position  the  instrument  with  enough  space  on  each  side  of  instrument  to  ensure  proper  air 
circulation. 

3.

 

ALWAYS  install the instrument  within  a  temperature  and  humidity  controlled  environment.  (Permissible  ambient 
temperature: +5°C ~ 

+35 °C, Relative humidity: ≤ 85%) 

4.

 

Before connecting the power, check the rated voltage.   

5.

 

Should not use unapproved rotors and accessories. 

Only use rotors from LaboGene, ApS with appropriate centrifugal tubes and suitable adaptors to embrace 
sample containers tightly enough inside rotors.

6.

Before operating the instrument, check if the rotor and the rotor lid are securely fastened.

 

Should operate the instrument with a rotor properly installed and secured to the motor shaft.   

7.

 

Mount the rotor on the motor shaft properly, check it with spinning manually. 

8.

 

Do not stop the rotor by touching with hand during the instrument is running. 

9.

 

Emergency door open should be performed only when spinning is completely stopped. 

10.

 

Should not exceed the rated speed or specific gravity. Samples whose density is greater than 1.2g/ml must have 
reduced maximum rotational speed to avoid rotor failure. 

11.

 

The  sample  content  should  not  exceed  80%  of  total  capacity  of  a  tube.  Otherwise,  it  would  cause  spillage  of 
sample fluid and even the tube breakage.   

12.

 

ALWAYS load the tubes symmetrically with evenly weighted samples to avoid rotor imbalance. If necessary, use 
the water blank to counterbalance the unpaired sample.   

13.

 

The operation speed should not exceed the highest value of the individual guaranteed g-forces of each centrifuge, 
rotor, bucket or adaptors and sample container, especially the guaranteed g-force of sample container should not 
be neglected.   

14.

 

The rotors should be cleaned and kept dry after every use for longer life and safety.   

15.

 

ALWAYS disconnect the power supply prior to maintenance care and service to avoid electrical shock.   

16.

ALWAYS use proven disinfection procedures after centrifuging bio-hazardous materials.

17.

 

Should not centrifuge flammable, toxic, radioactive, explosive, or corrosive materials.   

18.

When it is necessary to use toxic or radioactive materials or pathogenic micro-organisms which belong to the Risk 

Group II of WHO: “Laboratory Bio

-

safety Manual,” should follow national regulations.

Summary of Contents for ScanSpeed 1248

Page 1: ...Revision 01 Date 2013 05 07 Cat No 9 900 900 578 INSTRUCTION MANUAL MULTI PURPOSE HIGH SPEED CENTRIFUGE ScanSpeed 1248...

Page 2: ...01 Cat No 9 900 900 758 CopyRight 2013 Page 2 of 20 LaboGene ApS Industrivej 6 8 Vassinger d DK 3540 Lynge Denmark Phone No 45 39 40 25 66 Fax No 45 39 95 25 66 Mail info labogene com Internet www lab...

Page 3: ...07052013RS01 Cat No 9 900 900 758 CopyRight 2013 Page 3 of 20...

Page 4: ...07052013RS01 Cat No 9 900 900 758 CopyRight 2013 Page 4 of 20...

Page 5: ...Release 10 3 2 Rotor Coupling and Disassembling 11 3 3 Positioning of Sample Tubes 12 4 Operation 13 4 1 Key Functions of Control Panel 13 4 2 Setting RPM RCF 14 4 3 Setting Time 15 4 4 Acceleration...

Page 6: ...ning for rotor coupling Attention and warning for door opening and closing Attention and warning for correct way of sample balancing in the rotor Attention and warning for correct way of buckets posit...

Page 7: ...otor shaft 7 Mount the rotor on the motor shaft properly check it with spinning manually 8 Do not stop the rotor by touching with hand during the instrument is running 9 Emergency door open should be...

Page 8: ...control panel is dumb under the condition of complete stop of rotor running Never try to open or move the instrument if it is not completely stopped If the power input is more than 10 of the recommen...

Page 9: ...control Pulse timed 100 min or continuous RPM RCF conversion Yes Noise level dB 60 Acc Dec 9 10steps Program memory 100 Rotor Identification Automation Imbalance cutout Yes Display Blue LCD Automatic...

Page 10: ...shut When the door is opened the door LED turns on 3 2 Rotor Coupling and Disassembling The door is not opened while the instrument is running If the door is opened the instrument could not be operat...

Page 11: ...or manually to check if bucket swinging is free enough and ever If they do not swing freely apply the Lubricant grease to the linking area 3 2 2 Fixed Angle Rotor 1 Mount a proper rotor into the motor...

Page 12: ...s and do not exceed the speed beyond the tube s max g force For safety fill the sample for 70 80 in the tubes Correct Way of Sample Balancing Tube Usage If the number of samples is not in pair please...

Page 13: ...et the acceleration deceleration level from 1 to 9 steps 0 in deceleration step means natural deceleration Larger number means faster acceleration or deceleration PROG Use to save a set of setting val...

Page 14: ...0 rpm If you do not touch the button for 5 seconds the setting mode is cleared 3 Touch the Enter button to complete the setting Touch Enter to save the setting value 4 2 2 Setting RCF Speed setting un...

Page 15: ...sec from 1 sec If you do not touch the button for 5 seconds the setting mode is cleared 5 Touch the Enter button to complete the setting 4 4 Acceleration Deceleration Use the adjustment function of ac...

Page 16: ...setting mode is cleared Save up to 100 programs Program numbers from 00 to 99 4 Touch the Enter button to complete the saving 4 5 2 Recalling 1 To recall the saved program just touch the PROG button...

Page 17: ...ration is stopped 4 8 Emergency Door Open For emergency door opening you can use the Emergency Door Open Tool only when the instrument is completely stopped The door can be unlocked manually with the...

Page 18: ...h neutral detergents and keep dry 5 2 Chamber 1 Keep dry inside the chamber after every use 2 If the chamber is contaminated dip the cloth in neutral detergent and clean contaminated area 5 3 Shaft 1...

Page 19: ...rgency door tool manually for safety of sample Please refer to 4 8 Emergency Door Open Can t close the door Remove the dirt at the door latch and then close the door completely again If the door seems...

Page 20: ...ut of Power supply V Hz is 10 less than required power this message is appeared Shut off the power supply and then check the voltage of the Power supply V Hz Use AVR to provide proper power Error 5 Hi...

Page 21: ...irements EN 61010 2 020 Safety requirements for electrical equipment control and laboratory use Particular requirements for laboratory centrifuges EN 61000 6 1 Electromagnetic compatibility Generic im...

Reviews: