4-5.
Program Saving & Recalling
Action
4-5-1.
Saving
1.
Set parameters. (Refer to 4-2 ~ 4-4 )
2.
Touch the [PROG] button longer than 3 seconds.
Check the message of “PROGRAM SAVE: ##” on the
display window.
3.
Touch the
[▲▼] buttons to change input Program number.
If you do not touch
the [▲▼] button for 5 seconds, the
setting mode is cleared.
Save up to 100 programs. (Program numbers from 00 to
99)
4.
Touch the [Enter] button to complete the saving.
4-5-2.
Recalling
1.
To recall the saved program, just touch the [PROG] button shortly
(less than 1 sec.).
Check the message of “PROGRAM CALL: ##”, on the
display window.
2.
Touch the
[▲▼] buttons to select program number you want to
recall and then touch the [Enter] button.
If you do not touch
the [▲▼] button for 5 seconds, the
setting mode is cleared.
When you touch the [Enter] button, display window show the
saved setting parameters (RPM/RCF, TIME, TEMP).