4-2.
Setting RPM/RCF
Action
4-2-1.
Setting RPM
▶
Speed setting unit: 10 rpm or 100 rpm
1.
Touch the [RPM/RCF] button once.
RPM MODE is generated by pressing a [RPM/RCF] button
once.
RPM LED is flickering on the display window.
2.
Touch the [▲▼] buttons to change input value.
After keeping holding finger on the
[▲▼] buttons for 5
seconds, the unit of setting value is changed to 100 rpm
from 10 rpm.
If you do not touch
the [▲▼] button for 5 seconds, the
setting mode is cleared.
3.
Touch the [Enter] button to complete the setting.
Touch [Enter] to save the setting value.
4-2-2.
Setting RCF
▶
Speed setting unit: 1 rcf or 10 rcf
1.
Touch the [RPM/RCF] button once.
RPM MODE is generated by touching a [RPM/RCF] button
twice.
RPM LED is flickering at the display window.
2.
Touch the [
▲▼
] buttons to change input value.
After keeping holding finger on the
[▲▼] buttons for 5
seconds, the unit of setting value is changed to 10 rcf from 1
rcf.
If you do not touch the [
▲▼
] button for 5 seconds, the
setting mode is cleared.
3.
Touch the [Enter] button to complete the setting.