3-1.
Power On/Off and Door Release
Action
3-1-1.
Power On/Off
1.
After connecting the AC Power cord at the power socket on the right back of the
instrument, turn on the Earth Leakage Breaker Switch (1580 only).
Check the proper power.
2.
Turn on the instrument by pressing a switch on the right side of the instrument.
The default values are Max. rpm, 10 min, ACC 4, DEC 4.
3-1-2.
Door Release
1.
For opening the door, touch the [DOOR] button.
Should touch the [DOOR] button when the door is closed (Door
LED shows off).
Close the door until hearing clank shut.
When the door is opened, the door LED turns on.
3-2.
Rotor Coupling and Disassembling
☞
The door is not opened while the instrument is running.
If the door is opened, the instrument could not be operated even with pressing the [Start] button.
For operational safety, this instrument has the automatic rotor recognition function.
When you supply the power,
"Searching Rotor" /“Change to Rotor ID” will be appeared. If the rotor
is absent, the “Error 9” will be appeared. This message will be
cleared after rotor coupling and
running.
Power Failure: If there is any power failure during operation, door is not ope
ned with ‘Door’ button.
Door can be opened only when the operation is completely stopped and the power is on again. If
you want to open the door at the power failure, please refer to ‘4
-8
. Emergency Door Open’.