6.
Trouble Shooting
6-1.
Check List
Symptom
Check List
Power failure
Connect the AC Power cord and make sure that the line is completely connected
between the instrument and power outlet. Check the power switch is turned on.
(Please refer to 3-1. Power On/Off and Door Release)
Can’t be started
If the door is not closed
completely, the instrument can’t run.
Check the Door LED on the display window and close the door completely.
Can’t open the door
If the power is out, check the main fuse for the laboratory to supply the power. If it
is not solved in shortly, open the door with emergency door tool manually for safety
of sample. (Please refer to 4-8. Emergency Door Open)
Can’t close the door
Remove the dirt at the door latch and then close the door completely again.
If the door seems not being closed by mechanical reason, please contact our
service team.
Noise and vibration during running
Please check the balanced status of both the table and the instrument.
Please re-check the coupling status of the following three matches to minimize the
noise
1.
the balanced way of coupling of the rotor into the motor shaft
2.
the completeness of fixing of the Rotor Locking Nut on the rotor
3.
the matching status of Rotor Lid with the rotor
(Please refer to 3-2. Rotor Coupling and Disassembling)
Check balances of samples in the rotor. (Please refer to 3-3. Positioning of Sample
Tubes) and load the same weight of samples symmetrically.