INSTRUCTION MANUAL
MULTI-PURPOSE, HIGH SPEED CENTRIFUGE
ScanSpeed 1248
Page 1: ...Revision 01 Date 2013 05 07 Cat No 9 900 900 578 INSTRUCTION MANUAL MULTI PURPOSE HIGH SPEED CENTRIFUGE ScanSpeed 1248...
Page 2: ...01 Cat No 9 900 900 758 CopyRight 2013 Page 2 of 20 LaboGene ApS Industrivej 6 8 Vassinger d DK 3540 Lynge Denmark Phone No 45 39 40 25 66 Fax No 45 39 95 25 66 Mail info labogene com Internet www lab...
Page 3: ...07052013RS01 Cat No 9 900 900 758 CopyRight 2013 Page 3 of 20...
Page 4: ...07052013RS01 Cat No 9 900 900 758 CopyRight 2013 Page 4 of 20...
Page 5: ...Release 10 3 2 Rotor Coupling and Disassembling 11 3 3 Positioning of Sample Tubes 12 4 Operation 13 4 1 Key Functions of Control Panel 13 4 2 Setting RPM RCF 14 4 3 Setting Time 15 4 4 Acceleration...
Page 6: ...ning for rotor coupling Attention and warning for door opening and closing Attention and warning for correct way of sample balancing in the rotor Attention and warning for correct way of buckets posit...
Page 7: ...otor shaft 7 Mount the rotor on the motor shaft properly check it with spinning manually 8 Do not stop the rotor by touching with hand during the instrument is running 9 Emergency door open should be...
Page 8: ...control panel is dumb under the condition of complete stop of rotor running Never try to open or move the instrument if it is not completely stopped If the power input is more than 10 of the recommen...
Page 9: ...control Pulse timed 100 min or continuous RPM RCF conversion Yes Noise level dB 60 Acc Dec 9 10steps Program memory 100 Rotor Identification Automation Imbalance cutout Yes Display Blue LCD Automatic...
Page 10: ...shut When the door is opened the door LED turns on 3 2 Rotor Coupling and Disassembling The door is not opened while the instrument is running If the door is opened the instrument could not be operat...
Page 11: ...or manually to check if bucket swinging is free enough and ever If they do not swing freely apply the Lubricant grease to the linking area 3 2 2 Fixed Angle Rotor 1 Mount a proper rotor into the motor...
Page 12: ...s and do not exceed the speed beyond the tube s max g force For safety fill the sample for 70 80 in the tubes Correct Way of Sample Balancing Tube Usage If the number of samples is not in pair please...
Page 13: ...et the acceleration deceleration level from 1 to 9 steps 0 in deceleration step means natural deceleration Larger number means faster acceleration or deceleration PROG Use to save a set of setting val...
Page 14: ...0 rpm If you do not touch the button for 5 seconds the setting mode is cleared 3 Touch the Enter button to complete the setting Touch Enter to save the setting value 4 2 2 Setting RCF Speed setting un...
Page 15: ...sec from 1 sec If you do not touch the button for 5 seconds the setting mode is cleared 5 Touch the Enter button to complete the setting 4 4 Acceleration Deceleration Use the adjustment function of ac...
Page 16: ...setting mode is cleared Save up to 100 programs Program numbers from 00 to 99 4 Touch the Enter button to complete the saving 4 5 2 Recalling 1 To recall the saved program just touch the PROG button...
Page 17: ...ration is stopped 4 8 Emergency Door Open For emergency door opening you can use the Emergency Door Open Tool only when the instrument is completely stopped The door can be unlocked manually with the...
Page 18: ...h neutral detergents and keep dry 5 2 Chamber 1 Keep dry inside the chamber after every use 2 If the chamber is contaminated dip the cloth in neutral detergent and clean contaminated area 5 3 Shaft 1...
Page 19: ...rgency door tool manually for safety of sample Please refer to 4 8 Emergency Door Open Can t close the door Remove the dirt at the door latch and then close the door completely again If the door seems...
Page 20: ...ut of Power supply V Hz is 10 less than required power this message is appeared Shut off the power supply and then check the voltage of the Power supply V Hz Use AVR to provide proper power Error 5 Hi...
Page 21: ...irements EN 61010 2 020 Safety requirements for electrical equipment control and laboratory use Particular requirements for laboratory centrifuges EN 61000 6 1 Electromagnetic compatibility Generic im...