The packing lists should be referred to as soon as the units are received to
ensure that all components have been included. The unit should be checked
for damage when received. If damaged, contact carrier and save the box for
inspection. Please contact Labnet International if there are any problems or
missing items.
Usage Guidance and restrictions:
• Maximum altitude 2,000m.
• Temperature range between 4°C and 60°C.
• Maximum relative humidity 80% for temperatures up to 31°C decreasing linearly to
50% relative humidity at 40°C.
• Not for outdoor use.
This apparatus is rated POLLUTION DEGREE 2 in accordance with IEC 664.
POLLUTION DEGREE 2, states that: “Normally only non-conductive pollution
occurs. Occasionally, however, a temporary conductivity caused by condensation
must be expected.”
Care and Maintenance:
Cleaning ENDURO VE10 System
Units are best cleaned using warm water and a mild detergent.
Water at
temperatures above 60°
C can cause damage to the unit and components.
The
tank should be thoroughly rinsed with warm water and distilled water to prevent build
up of salts but care should be taken not to damage the enclosed electrode.
Vigorous cleaning is not necessary or advised.
Air drying is recommended before
use.
The units should never come into contact with the following cleaning agents,
these will cause irreversible and cumulative damage:-