10
them to medium wavelength (312 nm) UV light. Because the gel
casting tray is UV transmittant, the gel does not need to be
removed from the tray before viewing. Place the gel casting tray
containing the gel on the filter surface of a UV Transilluminator for
convenient viewing.
3. Sample banding patterns may be documented by
autoradiography
I. Trouble shooting guide
Problem
Cause
Solution
The LCD
screen is blank
AC power cord is not
connected
Check AC power cord connections at
both ends. Use the correct cords.
The power switch is not on Toggle the power switch
Operation stops
with alarm: The
screen displays
“
LOAD
”
Electrophoresis tank is not
connected to the power
supply or
there is a broken circuit in
the electrophoresis cell
Check the connections to the power
supply and on your electrophoresis cell
to make sure the connection is intact;
check condition of wires in
electrophoresis unit. Close the circuit by
reconnecting the cables. Press
RUN/PAUSE
to restart the run.
Buffer concentration
incorrect
Replace bufffer
Operation stops
with alarm:
Display shows
“
Lid
”
Lid was removed during a
run
• Verify that the lid is properly seated
• Verify the all connections are attached
correctly
• press the RUN/PAUSE button to
restart
Other error
• Turn off the power, disconnect the
power cord from the outlet, and contact
Technical Service