8
samples.
4. Load the samples into the wells with a micropipette or similar
device taking care not to puncture the bottom of the wells or load
the sample onto the top of the gel. .
F.
Electrical Connections to the Safety Lid and
The ENDURO Gel XL can only be operated with the safety lid in
place. Electrical current is supplied through the tank electrodes to
the power supply by placing the lid on the tank the circuit is
completed. A simple gravity connector in the cover ensures a
complete current path, yet allows the lid to be removed from the unit
without disturbing the loaded samples.
1. Make sure the power supply is turned off
2. Plug the male ends of the black (-) and red (+) electrodes into the
jacks on the side of the power supply. .
3. After the samples have been loaded into the gel, place the lid
over the unit so that the lid covers align with the tank.
4. Set the lid straight down so that the lid rests squarely on the tank,
connection is inside end of the lid which engages the power
supply.
5. Plug the Power Supply into a wall outlet.
Ensure an approved power cord that satisfies your regional
voltage standard is used.
Input voltage is automatically detected by the system. A
transformer is not necessary in Europe and any other region
where the standard voltage is higher than 100V.
6.
Set the timer. Increase or decrease the value with the Up and
Down buttons. Timer can be set between 1min – 99 hours. Set “--
:--” for continuous operation.
7.
Select the required output voltage up to 150 volt or 400 mA.
8.
Press the start/stop button to start the run.
To Pause a run and change parameters
.
1. To pause the run press the Run/Pause button once. During the
pause mode the voltage amperage or time can be changed by
highlighting the function and using the arrow keys then pressing
the mode key. Once the changes have been made the start
button can be pressed to resume the run.
2. To stop the run press the run/pause button for 3 seconds. Stop
will appear on the display.
CAUTION:
Do not jar or bump the gel box once the lid is place. The
safety switch is operated by a hall-effect sensor which relies
on a lid mounted magnet. Moving the gel box can move the
lid and cause the unit to pause until the lid is put back into
position.