Chapter 3: Getting Started
Product Service 1-800-522-7658
7
Unpacking the Enclosure
Carefully remove the shrink-wrap or carton on the enclosure and inspect it for
damage that may have occurred in transit. If damaged, notify the delivery carrier
immediately and retain the entire shipment intact for inspection by the carrier.
DO NOT RETURN GOODS WITHOUT THE PRIOR
AUTHORIZATION OF LABCONCO. UNAUTHORIZED
RETURNS WILL NOT BE ACCEPTED.
IF ENCLOSURE WAS DAMAGED IN TRANSIT, YOU MUST
FILE A CLAIM DIRECTLY WITH THE FREIGHT CARRIER.
LABCONCO CORPORATION AND ITS DEALERS ARE NOT
RESPONSIBLE FOR SHIPPING DAMAGES.
Do not discard the packing material until you have checked all of the components
and tested the enclosure. We recommend that you do not remove the enclosure
from its shipping container until it is ready to be placed into its final location.
Move the unit by using a floor jack.
Do not move the enclosure by tilting it onto a hand truck.
Installing the Enclosure on a Supporting
Structure and Work Surface
When installing the enclosure onto a chemically-resistant work surface or
benchtop, ensure that the structure can safely support the combined weight of the
enclosure and any related equipment. The work surface should be at least as wide
and deep as the non-ventilated enclosure to properly support it. The front of the
non-ventilated enclosure should be aligned within 1/8" of the front of the work
surface. Mounting holes are provided in the Labconco accessory work surfaces to
secure the enclosure.
Work Surface Specifications
The work surface should be smooth, rigid, and durable, such as a chemically-
resistant epoxy resin. The surface should be
non-porous and resistant to the materials used in conjunction with the Purifier PCR
or Purifier Tissue Culture Enclosures. The work surface should also contain a
dished recessed area for containing primary spills.
The United
States
Interstate
Commerce
Commission
rules require
that claims be
filed with the
delivery
carrier within
fifteen (15)
days of
delivery.
!