Chapter 3: Getting Started
Product Service 1-800-522-7658
8
8
Work Surface and Non-Ventilated Enclosure Installation
1.
Level the base cabinets and the work surface. Work surface should be
placed flush with the front of the base cabinet as shown in Figure 3-1.
2.
Position the work surface in its intended location and with the front of the
work surface, (wide flange), towards you.
3.
Secure the work surface to the base cabinet with a structural adhesive or
silicone sealant.
4.
Insert the supplied mounting screws in the four holes. Allow a minimum
of 1/8" clearance under the head of the screw for positioning the enclosure.
5.
Place the enclosure on the work surface and slide the rear flange and lower
front bracket flanges under the mounting screw heads.
6.
Tighten the four screws to complete the installation.
Figure 3-1