39
Product Service 1-800-522-7658
Chapter 8: Troubleshooting
PROBLEM
CAUSE
CORRECTIVE ACTION
Vertical sash frame is
distorted.
Place horizontal glass symmetrically
and pull sash down to air foil.
Straighten damaged frame.
Cable is frayed or has
slipped off the
pulleys.
Re-install, cable must be replaced
immediately if damaged.
Electrical duplex
outlets no longer
have power
Wires not connected
or faulty duplex.
Check wire connection or replace
duplex.
Circuit
breakers
tripped in building
electrical supply.
Reset circuit breakers.
Service valves no
longer operate
Faulty building
supply.
Inspect building supply shut off valves
and appropriate pressures below 40
PSI.
Valve no longer
operates.
Replace valve and check for leaks.
Supply line or outlet
line has leaks.
Inspect line for leaks and fix any
leaking plumbing connections.
Summary of Contents for Protector XStream 9840400
Page 23: ...18 Product Service 1 800 522 7658 Chapter 3 Getting Started...
Page 35: ...30 Product Service 1 800 522 7658 Chapter 5 Using Your Protector Fume Hood...
Page 45: ...40 Product Service 1 800 522 7658 Chapter 8 Troubleshooting...
Page 49: ...44 Product Service 1 800 522 7658 Appendix A Protector Components...
Page 51: ...46 Product Service 1 800 522 7658 Appendix B Protector Dimensions...
Page 53: ...48 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Page 54: ...49 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Page 55: ...50 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Page 56: ...51 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Page 57: ...52 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Page 61: ...56 Product Service 1 800 522 7658 Appendix D References...