29
Product Service 1-800-522-7658
Chapter 5: Using Your Protector Fume Hood
Work Techniques
•
Keep all materials at least 6 inches inside of the sash, and
perform all contaminated operations as far to the rear of the
work area as possible.
•
Segregate all clean and contaminated materials in the work
area.
•
Avoid using techniques or procedures that disrupt the
airflow patterns of the hood.
Final Purging
•
Upon completion of work, the hood should be allowed to
operate for two to three minutes undisturbed, to purge
airborne contaminants from the work area before shutting
down blower.
Unloading Materials and Equipment
•
Objects in contact with contaminated material should be
surface decontaminated before removal from the hood.
•
All open trays or containers should be covered before being
removed from the hood.
Shutdown
•
Turn off the fluorescent light and hood blower, then close
the sash.
Summary of Contents for Protector XStream 9840400
Page 23: ...18 Product Service 1 800 522 7658 Chapter 3 Getting Started...
Page 35: ...30 Product Service 1 800 522 7658 Chapter 5 Using Your Protector Fume Hood...
Page 45: ...40 Product Service 1 800 522 7658 Chapter 8 Troubleshooting...
Page 49: ...44 Product Service 1 800 522 7658 Appendix A Protector Components...
Page 51: ...46 Product Service 1 800 522 7658 Appendix B Protector Dimensions...
Page 53: ...48 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Page 54: ...49 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Page 55: ...50 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Page 56: ...51 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Page 57: ...52 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Page 61: ...56 Product Service 1 800 522 7658 Appendix D References...