14
Product Service 1-800-522-7658
Chapter 3: Getting Started
for minimum static pressure loss with proper
transport velocities away from the hood. Should
higher transport velocities be required, simply
install a reducer after the hood exhaust outlet.
Consult Labconco Customer Service should you
require help sizing your blower for the exhaust
volume and total system static pressure loss.
The selected exhaust duct material should match the
hood procedures and chemicals used to ensure
compatibility.
Connecting the Electrical Supply
Source to the Protector XStream
Fume Hood
Prior to connecting the Protector XStream to any electrical wiring
to the fume hood structure, refer to the hood identification plate for
the proper electrical requirements of your specific model.
The identification plate, model number, serial number, and
electrical connection boxes are accessible from the front of the
fume hood by removing the front panel.
The Protector XStream Hood is wired for 115 Volt, 60 Hz, 20
Amp or 230Volt, 50 Hz, 10 Amp electrical service. Check the I.D.
plate behind the front panel for voltage verification. The number
of circuits varies depending on the model. All of the electrical
connections are terminated at the single point internal junction box
for hook-up by a qualified electrician. The single point internal
junction box is used for the connection of the lights, blower, and
duplex outlets. Refer to the wiring diagram for your Protector
XStream in
Appendix C: Protector XStream Fume Hood
Specifications.
WARNING
:
The building electrical supply system
for Protector XStream Hoods should include
overload protection. A switch or circuit breaker
should be in close proximity to the equipment and
within easy reach of the operator. The switch or
circuit breaker is to be marked as the disconnecting
device for the equipment. Consult the NEC-2002 for
proper installation.
!
!
Summary of Contents for Protector XStream 9840400
Page 23: ...18 Product Service 1 800 522 7658 Chapter 3 Getting Started...
Page 35: ...30 Product Service 1 800 522 7658 Chapter 5 Using Your Protector Fume Hood...
Page 45: ...40 Product Service 1 800 522 7658 Chapter 8 Troubleshooting...
Page 49: ...44 Product Service 1 800 522 7658 Appendix A Protector Components...
Page 51: ...46 Product Service 1 800 522 7658 Appendix B Protector Dimensions...
Page 53: ...48 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Page 54: ...49 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Page 55: ...50 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Page 56: ...51 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Page 57: ...52 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Page 61: ...56 Product Service 1 800 522 7658 Appendix D References...