Chapter 2: Prerequisites
by a qualified electrician. Refer to
Chapter 3: Getting Started
and
Appendix
C
: Protector XStream Specifications
for the wiring
diagram for proper electrical installation.
Service Line Requirements
All service lines to the laboratory fume hood should be ¼ inch
outside diameter, copper (brass for natural gas), and equipped with
an easily accessible shut-off valve, should disconnection be
required. If the service line pressure exceeds 40 PSI, it must be
equipped with a pressure regulator to reduce the line pressure.
Please check with local codes for other requirements.
Space Requirements
The dimensions for the different models are shown in
Appendix B:
Protector XStream Dimensions
.
Summary of Contents for Protector XStream 9840400
Page 23: ...18 Product Service 1 800 522 7658 Chapter 3 Getting Started...
Page 35: ...30 Product Service 1 800 522 7658 Chapter 5 Using Your Protector Fume Hood...
Page 45: ...40 Product Service 1 800 522 7658 Chapter 8 Troubleshooting...
Page 49: ...44 Product Service 1 800 522 7658 Appendix A Protector Components...
Page 51: ...46 Product Service 1 800 522 7658 Appendix B Protector Dimensions...
Page 53: ...48 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Page 54: ...49 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Page 55: ...50 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Page 56: ...51 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Page 57: ...52 Product Service 1 800 522 7658 Appendix C Protector Hood Specifications...
Page 61: ...56 Product Service 1 800 522 7658 Appendix D References...